Skip to content

Global shipping available. Please request a quote to receive customized pricing for your region.

Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485 Default Title Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485 Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO850 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI480 Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO850 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI480 Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485 Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485 Default Title Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485 Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO850 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI480 Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO850 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI480 Portable Optical Dissolved Oxygen Meter Apera Instruments
Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485 Portable Optical Dissolved Oxygen Meter Apera Instruments

Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485

$1,350.00

    Measure DO, account for salinity and keep up to 500 timestamped records.

    • 10-ft nominal DO803 optical probe plus 2301-3M salinity electrode
    • Automatic temperature, pressure and selectable automatic salinity compensation
    • 500 stored groups with date/time and manual or timed saving
    • USB output using the documented legacy DO8500 PC-Link workflow

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

Preparing Quote Cart. Your shopping cart stays unchanged.

APERA INSTRUMENTS / DO8500 / AI485

Optical DO with salinity compensation and a record of the round

DO8500 adds timestamped storage and automatic salinity compensation to a portable optical DO workflow. The AI485 kit includes the DO803 optical probe with nominal 10-ft cable, a separate salinity electrode and a clip for measuring together. It is designed for general-water sampling where salinity changes and saved results both matter.

Choose DO8500 when manual or timed storage and a salinity input measured by the included electrode are useful. The salinity electrode supports DO compensation; this is not a general pH/EC multiparameter kit.

DO8500 optical DO meter with nominal 10-ft cable and supplied probes
AI485: DO8500 with DO803 and separate salinity electrode joined by the clip. Exact 10-ft kit page photograph; cable length is established by SKU and manual, not inferred from the coiled cable.
Which meter kit fits my measurement?Click or tap to compare all six measurement configurationsClick or tap here to close the guide

Each card opens that model page. Match the supplied electrode and documented sample method. pH, conductivity and optical dissolved oxygen are different measurements. Match the parameter and meter format before comparing accessories; optional items are not supplied unless the exact kit list says so.

Choose the electrode for the sample.

Optical sensing without oxygen consumption

A blue LED excites the luminescent layer; a photodiode detects its red optical response. The probe does not consume oxygen, so measurement does not require continuous sample movement. Still remove bubbles, inspect the cap and allow temperatures to equalize. Optical design is not a guarantee of immediate readings in every sample.

Salinity compensation has its own electrode

The included 2301-3M conductivity electrode supplies the salinity input used for automatic DO compensation. Connect it and choose automatic mode; use the combined clip to measure with the DO probe. Manual entry from 0 to 45 ppt remains available. Do not assume the optical probe alone measures salinity.

Read status before saving

The LCD distinguishes a stable reading, HOLD, M+ saving, RM recall, timed storage and USB communication. The manual shows salinity when its probe is connected and salinity exceeds 1 ppt. HOLD is a reading lock, separate from the commands that actually store records.

Field handling with a serviceable sensor

The IP57 instrument has a backlight and folding stand, and travels in a carrying case. The DO803 integrates temperature sensing and uses a replaceable optical cap. Protect connectors and seals, and keep the cable clear of sharp edges. The ingress rating does not make every accessory or connected computer waterproof.

Know what this configuration supplies.

Included item Order contents
DO8500 portable optical DO meter One instrument, exact US kit SKU AI485.
DO803 optical DO probe One temperature-sensing optical probe with fitted cap and nominal 10-ft cable.
Calibration/storage sleeve and sponges One DO8031 sleeve; four spare water-storage sponges.
2301-3M salinity electrode and combined clip One conductivity electrode for salinity compensation and one clip for paired measurement.
PC-Link flash drive and USB cable One of each listed in the exact manual. Host compatibility requires confirmation; computer is not included.
Screwdriver, case and instructions One small screwdriver, one carrying case and one manual. The meter uses three AA batteries; the linked packing list does not separately enumerate batteries or salinity calibration solution.
DO8500 kit in an open carrying case
Manufacturer photograph from the exact 10-ft kit page. Use the itemized kit list for included components, rather than counting hidden pieces in the case.

Match the sample and preparation method.

Brackish-water and marine checks

Automatic salinity compensation reduces repeated manual entries when moving among waters of different salinity. It depends on the connected, functioning salinity electrode and the selected mode. Record sampling depth and temperature so comparisons reflect a consistent procedure.

Treatment-process sampling rounds

Store readings with date and time at successive locations, then preserve location identifiers in your own records. Timed recording can follow a defined sampling interval; it does not by itself establish sample representativeness or replace an installed process sensor.

General-water laboratory comparisons

The stand and probe clip support repeated bench checks. Use compatible aqueous samples within 0 to 50°C. DO803 is not specified for strong acid, strong alkali or organic solutions; determine suitability before applying it to aggressive process liquids.

Original optical DO principle diagram showing blue excitation and red luminescence
Original manufacturer principle illustration: the blue LED excites the sensing layer and the photodiode detects red luminescence. This explains the design, not a test result or an exact immersion-depth drawing.

Prepare the DO and salinity paths separately

Connect the DO803 eight-pin plug, secure the connector and inspect the sensing cap. Allow about 30 seconds after power-on before operations. Check the clock if storing data. Select automatic or manual salinity compensation deliberately; automatic mode requires the salinity electrode to be connected, not merely a setting change.

The salinity electrode is factory calibrated. The manual calls for recalibration when it is replaced or has not been used for a long time. In automatic mode use the specified 12.88 mS/cm conductivity standard in the salinity calibration menu, allow the reading to stabilize, then confirm. This standard is for the salinity electrode, not the optical DO saturated-air point.

For DO calibration, use the storage sleeve with a damp, non-dripping sponge. Keep water droplets off the sensor cap and wait five to ten minutes for humidity and temperature to stabilize. Confirm the saturated-air calibration only after the stability icon remains visible. The manual recommends daily saturated-air calibration for optimal accuracy, despite promotional suggestions of infrequent calibration.

Zero calibration is factory performed and normally unnecessary on initial use. It is specified after probe or cap replacement, prolonged non-use, or for demanding work at 0 to 2.0 ppm. Follow the manual’s fresh sodium-sulfite solution procedure. For measurements remove the sleeve, immerse above the marked line, briefly dislodge bubbles and wait for stability. Full equilibration can take about one to three minutes; the 30-second specification is only 90% response at 25°C.

Original numbered DO803 sensor and calibration sleeve assembly diagram
Original manual diagram. Legend: 1 sleeve bottom cover; 2 water-storage sponge; 3 sensor cap; 4 calibration/storage sleeve; 5 locking cap; 6 optical probe; 7 O-ring. For saturated-air calibration keep the sponge damp, not dripping, and keep droplets off the sensing face.
Original DO803 immersion-line drawing
Original manual depth guide: during sampling the water must be above the marked immersion line. Remove the storage sleeve for measurement. This is not the humid-air calibration arrangement.
Original three-step DO8500 salinity calibration screen sequence
Original screens from manual Figure 13, arranged vertically for small displays. Select automatic salinity mode, use 12.88 mS/cm conductivity standard for the salinity electrode and confirm after stability. These are salinity calibration screens, not the optical DO saturated-air calibration.

Understand the sensing end and connections.

DO803 optical probe sensing end with stainless-steel cap
DO803 sensing end, shown without its storage sleeve. Protect the optical coating from scratching; do not exceed 50°C.

Keep the cap humid, clean and protected

The optical cap needs humid storage, not pH-electrode KCl solution. Keep the DO8031 sleeve sponge damp with clean water, not dripping, and close the locking cap. The sensing cap should sit in humid air rather than a pool of water inside the sleeve. Replace a dirty or moldy sponge. For storage beyond 30 days, the manual permits clean-water storage or checking the sleeved sponge every 30 days.

If the sponge has dried or the probe has spent over eight hours in dry air, rehydrate before measuring. The detailed maintenance instructions specify 24 hours in clean water at room temperature, or 48 to 72 hours in cold water. During measurement, remove the sleeve and immerse above the marked immersion line. Do not confuse that sampling depth with the humid-air calibration arrangement.

Protect the cap coating from scratches and temperatures above 50°C. Clean gently with a soft clean cloth; do not use alcohol or organic solvents. The manual describes 3% hydrogen peroxide for 15 to 30 minutes when disinfection is needed, followed by a clean-water rinse. Do not routinely remove or swap sensor caps; keep the interior dry and uncontaminated.

DO8032 is the replacement luminescent cap. Its advertised 8,000-hour service figure is a design estimate, not a guaranteed interval or warranty duration. Replacement requires the correct cap calibration coefficients. The current replacement page describes a probe-connected data box and installation-program flash drive, while the meter manual refers to cap-specific calibration codes. Follow the supplied replacement instructions; neither source establishes a QR-code or RFID procedure.

DO8031 storage and calibration sleeve with sponge at its base
The DO8031 sleeve creates humid air around the sensing cap. Dampen the sponge with clean water; do not fill the sleeve so the sensing cap sits in water. Inspect and replace dirty or moldy sponges.
Original DO803 exploded photo with luminescent cap removed
Original DO8032 service illustration, not normal storage: remove the cap only for appropriate service, keep the interior dry and follow the replacement instructions. No conventional membrane or electrolyte refill is needed.
DO803 replacement-cap data box connected between probe and a computer
Replacement-cap programming illustration from the DO8032 accessory page. It shows a data box connected to the probe and a computer; it is not ordinary meter data export. Computer and cap-service equipment are not listed as included in the meter kit. Confirm the supplied cap procedure and host requirements.
Original DO8500 LCD guide with numbered status icons
Original LCD legend: 1 mode; 2 reading; 3 timed storage; 4 date/time; 5 date/time units; 6 measurement units; 7 temperature units; 8 calibration units; 9 storage number; 10 M+ save and RM recall; 11 temperature; 12 HOLD; 13 ATC/MTC; 14 stability; 15 USB; 16 low battery. This diagram shows possible segments, not a sample reading.
DO8500 display with salinity value highlighted
Original manufacturer DO8500 example: the lower-right salinity value is highlighted. Automatic salinity compensation requires the included separate electrode and automatic mode; the optical probe alone does not measure salinity.
DO8500 pressure compensation menu example
Original DO8500 P3.5 pressure-calibration display. Automatic barometric compensation operates over 60 to 120 kPa. The example number is not a measured value for your location.
DO8500 supported on its folding rear stand
Original exact DO8500 feature-panel photograph showing the folding stand. Secure the instrument and cables on a stable surface; the stand does not secure the probes in a sample.

Read the display. Check the measurement mode.

Use the physical controls for measurement, calibration and the documented record functions. Check units, temperature, calibration indicators and the selected parameter before assigning a result to a sample. Screen values in photographs are illustrative, not sample results. The model-specific record and output workflow is explained below.

Follow the exact parameter-specific calibration procedure, using the required standard, solution or calibration atmosphere. Temperature compensation does not make different sample temperatures, preparation methods or chemistry equivalent. Follow the model-specific meter and electrode instructions before extending the method to a new sample.

500 records, with a model-specific legacy USB route

DO8500 stores up to 500 groups with date and time. Use manual saving for individual samples or set the automatic timing interval for a series. Manual saving is unavailable while timed-storage mode is active. Check the clock before collecting a batch, review records with RM and export before clearing the complete memory.

The exact manual documents DO8500 PC-Link, a Windows driver and COM-port connection, Download and Export to a Microsoft Excel file. In connected real-time operation, readings are sent to the computer rather than retained in meter memory. Keep the computer session and exported file under your own record-retention procedure.

This is the legacy DO8500 workflow, not evidence of compatibility with 9500 software, 480-series software or a browser-based PC-Link service. Verify the delivered meter revision and host-system requirements before purchase if USB use is essential. No software installation, startup or compatibility test is implied, and the kit is not represented as compliance-certified data management.

DO8500 connected to a computer by USB
Original DO8500 USB illustration. Up to 500 timestamped groups are supported; the exact manual describes legacy PC-Link. Computer not included. This is not evidence of browser support or validated compliance software.
Original DO8500 legacy PC-Link software screen
Legacy manual interface, not a compatibility test. Legend: 1 stored values; 2 mode and serial number; 3 parameter information; 4 port and command controls. Download transfers saved meter records; Export writes an Excel file. Connected live records go to the PC rather than meter memory.

Complement your water-quality workflow.

Related workflow: Apera TN420 Portable White-Light Turbidity Meter. Turbidity and dissolved oxygen describe different water-quality properties; this link is a separate workflow, not a compatible accessory or interchangeable measurement. Compare the exact supplied sensors, measurement parameter, accessories and data features rather than assuming similar model names mean identical kits.

DO8500 buyer questions

Which exact configuration is this?

AI485 is the nominal 10-ft DO8500 kit, with DO803 and a separate salinity electrode. The manual calls the optical cable 3 m. This is not the AI48533 33-ft configuration.

What are the DO and temperature ranges?

The meter covers 0 to 20.00 mg/L or ppm, 0 to 200.0% saturation and 0 to 50°C. Salinity compensation covers 0 to 45 ppt; automatic pressure compensation covers 60 to 120 kPa.

Is the salinity sensor included or optional?

Included. The manual identifies the supplied electrode as 2301-3M and also lists a combined electrode clip. Automatic compensation requires that sensor to be connected and automatic mode selected.

Does it measure pH and general conductivity?

No pH measurement is specified. The conductivity electrode supports salinity compensation of DO; do not treat this as a general-purpose pH/EC kit.

How much data can it save?

Up to 500 timestamped groups, using manual or timed storage. HOLD alone does not save a record. Connected real-time PC-Link acquisition is described as storing on the computer, not in the meter.

Will USB work in a modern browser?

Browser support is not established by the exact DO8500 sources. Its manual documents legacy Windows PC-Link and a COM-port/driver workflow. Confirm current host compatibility for your meter revision.

Does the cap last a guaranteed 8,000 hours?

No. That is a service-life design estimate, not a guaranteed duration. Coating damage, unsuitable cleaning and dry storage can affect performance. Replacement also requires the correct calibration coefficients.

What warranty applies?

The current exact kit page states three years for the instrument, two years for the DO803 probe excluding its cap, and one year for the DO8032 cap, from purchase. Coverage concerns material or manufacturing defects; improper use, normal wear, accidental damage, storage, transportation and unauthorized repair are excluded. The service-hour estimate is not a warranty. The linked DO8500 manual instead dates cap coverage from delivery; confirm the applicable terms for your purchase.

Confirm the exact configuration before ordering.

Review DO8500 AI485 against the sample, supplied electrode and documented method. Return to the live price and ordering controls when the configuration fits your work.

Back to price and ordering ↑
    SpecificationDetail
    Model / kit SKUDO8500 / AI485
    MeasurementLuminescent optical dissolved oxygen and temperature
    DO concentration range0 to 20.00 mg/L (ppm)
    Saturation range0 to 200.0%
    Selectable resolution0.01 or 0.1 mg/L (ppm); 0.1 or 1%
    DO concentration accuracy±2% of reading or ±0.2 mg/L, whichever is greater
    Saturation accuracy±2% of reading or ±2% saturation, whichever is greater
    Response specification90% response within 30 seconds at 25°C; full stabilization takes longer
    CalibrationSaturated-air point and zero-oxygen point
    Temperature range / resolution / accuracy0 to 50°C / 0.1°C / ±0.5°C
    Temperature compensationAutomatic, 0 to 50°C
    Barometric compensationAutomatic, 60 to 120 kPa
    Power / ingress ratingThree AA batteries / IP57
    Probe / cableDO803 with built-in temperature sensor; 10-ft nominal cable (manual: 3 m)
    Salinity compensationManual or automatic with included 2301-3M, 0 to 45 ppt
    Data handling500 timestamped groups; USB; manual/timed storage
    Limited warrantyInstrument 3 years; DO803 excluding cap 2 years; DO8032 cap 1 year; exclusions apply

    Ordering Information

    Ordering detailConfiguration
    Manufacturer catalogAI485
    Blue Tiger SKUBTS-AI485
    Order unit1 exact listed meter-kit configuration
    Order contentsDO8500 portable optical DO meter: One instrument, exact US kit SKU AI485.; DO803 optical DO probe: One temperature-sensing optical probe with fitted cap and nominal 10-ft cable.; Calibration/storage sleeve and sponges: One DO8031 sleeve; four spare water-storage sponges.; 2301-3M salinity electrode and combined clip: One conductivity electrode for salinity compensation and one clip for paired measurement.; PC-Link flash drive and USB cable: One of each listed in the exact manual. Host compatibility requires confirmation; computer is not included.; Screwdriver, case and instructions: One small screwdriver, one carrying case and one manual. The meter uses three AA batteries; the linked packing list does not separately enumerate batteries or salinity calibration solution.

    Apera Manufacturer Limited Warranty

    The current exact kit page states three years for the instrument, two years for the DO803 probe excluding its cap, and one year for the DO8032 cap, from purchase. Coverage concerns material or manufacturing defects; improper use, normal wear, accidental damage, storage, transportation and unauthorized repair are excluded. The service-hour estimate is not a warranty. The linked DO8500 manual instead dates cap coverage from delivery; confirm the applicable terms for your purchase.

    Keep the exact product and purchase details when requesting service. Follow the stated model-specific warranty conditions and product instructions; the instrument and supplied sensors or consumables can have different coverage terms.

    DO8500 official documents

    Use the meter and electrode documents for their stated model and scope. Regional editions are labeled where supplied; software, probe and packing details can differ between revisions. Check the supplied instrument markings when sources describe different hardware revisions. Each card opens a verified manufacturer PDF in a new tab.