Swap probes, confirm stability and save selected readings in 400-set internal memory.
Preparing Quote Cart. Your shopping cart stays unchanged.
APERA INSTRUMENTS / SX736 / AI468
SX736 combines pH, conductivity and dissolved oxygen testing in a battery-powered carrying-case kit. Connect the appropriate supplied electrode, let the meter recognize its mode, and take one measurement at a time. Conductivity also supports TDS, salinity and resistivity displays; these are related outputs, not additional independent probes. Temperature accompanies supported electrode measurements.
Choose AI468 for sequential spot checks with replaceable 201T-S, 2301T-S and DO500 electrodes and on-meter recall. The kit includes polarographic DO500, not an optical oxygen sensor. An ORP electrode is not supplied. It is not a simultaneous multichannel logger or a wireless app-based tester.
Each card opens that model page. Match the supplied electrode and documented sample method. pH, ORP, conductivity and polarographic dissolved oxygen are different measurements. Match the parameter and meter format before comparing accessories; optional items are not supplied unless the exact kit list says so.
AI460
pH and temperature
201T-S pH/temperature probe; ORP electrode separate.
AI463
pH and ORP kit
Separate pH and ORP probes; one connection at a time.
AI477
pH and conductivity kit
Separate 201T-S and 2301T-S probes; sequential tests.
AI461
Polarographic dissolved oxygen
DO500 membrane probe; polarization and movement required.
AI465
pH and dissolved oxygen
Swap the pH and DO500 probes for sequential tests.
AI467
pH, ORP and conductivity
Three supplied sensor types; not simultaneous acquisition.
AI468
pH, conductivity and DO
Three probe workflows with distinct storage and calibration.
AI469
pH, ORP, conductivity and DO
Four probe workflows; one input and model-specific records.
Automatic recognition selects the measurement family for the connected probe. Use the keypad to choose conductivity, TDS, salinity or resistivity, and the desired resolution or units. This changes the displayed result, not the number of sensors connected. The backlit LCD includes temperature, stability, calibration and memory indicators.
The pH meter spans -2.00 to 19.99 pH, but the current US 201T-S probe page specifies 0 to 13 pH and 0 to 60°C. It is intended for general aqueous samples, not every solution within the meter display range. Older probe instructions show wider limits; use the conservative current probe specification when selecting an application. The current 201T-S listing excludes low-ion water, strong-base samples above pH 12 and highly corrosive liquids. Its probe guide also cautions against frequent testing below pH 2 or above pH 12 and prolonged exposure above 50°C; 0 to 60°C is not a continuous-use guarantee.
The supplied 2301T-S uses a K=1 platinum-black cell and a temperature sensor. Current US documentation rates the probe to 200 mS/cm; the meter display ends at 199.9 mS/cm. K=0.1 and K=10 settings accommodate other suitable cells, not extra electrodes in this kit. Verify high-range work against an appropriate standard.
Compare compatible water samples using consistent containers, immersion and temperature equilibration. Record the selected parameter and units with every result. Conductivity measures aggregate ionic response; calculated TDS is not a gravimetric solids test, and an ORP or pH reading alone cannot establish drinking-water safety.
The IP57-rated handheld format and carrying case suit spot-check routes where the operator can rinse probes between sites. Use the 400-set memory alongside a sample log. Protect the connector from contamination, keep spare solutions closed, and respect the stated 5 to 35°C instrument working environment.
For ponds, aquaculture or surface-water checks, allow time for DO polarization and sample flow. Do not treat a stationary oxygen probe as an unattended monitor. Use representative sampling and separately document pH and conductivity rather than implying all readings occurred at exactly the same instant.
Align the 8-pin plug notch with the single meter socket, push straight in and tighten only the metal locking ring. Do not twist the black rubber cable section. Rinse the selected electrode with distilled or deionized water before testing, remove excess water gently and avoid contaminating fresh standards.
For pH, use the selected USA, NIST or China buffer series and one to three recognized calibration points. The supplied bottles are pH 4.00, 7.00 and 10.01. The manual calls for three-point calibration with a new pH electrode. Match calibration and sample temperatures where practical; automatic temperature compensation corrects electrode response, not every chemical temperature effect.
For conductivity, calibrate in conductivity mode with one recognized standard close to the sample range. The kit supplies 1413 µS/cm, not every supported standard. Reference temperature choices are 18, 20 or 25°C; the linear coefficient adjusts from 0.00 to 9.99%/°C, default 2.00. Record settings before comparing batches.
For DO500, polarize for five minutes after power-on before calibration or testing. Use 30 minutes after replacing a membrane cap or inner solution. Calibrate at saturation in the sleeve with a moist, non-dripping sponge and wait for stability. Temperature and salinity compensation are automatic; barometric pressure is entered manually.
Routine DO calibration is saturation calibration. Optional zero-oxygen calibration follows the exact model manual and requires separately obtained supplies. Sodium sulfite and the 12.88 mS/cm standard for salinity-sensor recalibration are not listed kit supplies. DO502 is electrode electrolyte, not a calibration standard.
Keep the 201T-S glass bulb hydrated in its 3M KCl storage solution. Do not store a pH electrode in distilled or deionized water. By contrast, the 2301T-S conductivity cap uses clean distilled or deionized water. Rinse between samples and keep the electrical plug dry; never interchange these storage routines.
The low-conductivity nonlinear compensation function is described for readings below 10 µS/cm. It does not make the supplied pH electrode suitable for ultrapure water. For very low ionic strength pH samples, select an appropriate specialist electrode. Low-range conductivity may require a compatible K=0.1 cell and a controlled flow-cell method.
DO500 consumes oxygen during measurement. In still water, the manual requires continuous slow swirling above 5 cm/s, including while recording. Keep the temperature sensor submerged and the probe at 45 to 75 degrees to the surface. Inspect the membrane and electrolyte for damage or bubbles; use DO503 caps and DO502 refill rather than treating it as a no-refill optical probe.
The current 201T-S listing excludes low-ion water, strong-base samples above pH 12 and highly corrosive liquids. Its probe guide also cautions against frequent testing below pH 2 or above pH 12 and prolonged exposure above 50°C; 0 to 60°C is not a continuous-use guarantee.
DO500 is intended for general water solutions, not strong acidic, strong alkaline or organic solutions. Its probe operating temperature is 0 to 50°C, distinct from DO compensation at 0 to 45°C and meter operating conditions.
Use the physical controls for measurement, calibration and the documented record functions. Check units, temperature, calibration indicators and the selected parameter before assigning a result to a sample. Screen values in photographs are illustrative, not sample results. The model-specific record and output workflow is explained below.
Follow the exact parameter-specific calibration procedure, using the required standard, solution or calibration atmosphere. Temperature compensation does not make different sample temperatures, preparation methods or chemistry equivalent. Follow the model-specific meter and electrode instructions before extending the method to a new sample.
The exact SX736 manual documents 400 stored sets. A short memory-key press saves a stabilized result; a long press enters recall. Records contain a serial number, measurement value, unit and temperature. The stated capacity is not a promise of 400 records for every mode or simultaneous complete-parameter datasets.
M+ and RM identify saving and recall. The manual does not document a date/time clock, timed-interval storage, USB export, Bluetooth, browser export or an electronic GLP audit trail. Record sample identity and collection time separately. Deleting memory clears all saved records, so transcribe the required values before using that command.
Related workflow: Apera DO8500 Portable Optical Dissolved Oxygen Meter Kit, 10-ft Cable | AI485. The linked optical DO kit has a different sensor technology and maintenance workflow; its probe or cap is not interchangeable with DO500. It is a DO-workflow option, not a replacement for this kit’s other parameters. Compare the exact supplied sensors, measurement parameter, accessories and data features rather than assuming similar model names mean identical kits.
The supplied electrode models are 201T-S, 2301T-S and DO500. All connect through the single 8-pin input. DO500 is polarographic, with replaceable membrane and electrolyte. SX736 does not include 301Pt-S; do not assume the kit includes ORP hardware.
No. Change the connected electrode for a different measurement family. Recognition selects that probe’s mode; conductivity-derived TDS, salinity and resistivity are display choices, not extra independent sensors. Temperature can accompany supported measurements.
201T-S is AI1312; 2301T-S is AI1313; DO500 is AI4110. DO502 refill is AI4112 and DO503 membrane caps are AI4113. Choose the S-version 8-pin connection, not a T-F BNC/RCA substitute. The optional AI7103 adapter is not supplied.
Not without considering electrode limits and sample chemistry. The current 201T-S page narrows its range to 0 to 13 pH and 0 to 60°C; older manuals differ. The 2301T-S is rated 0.5 µS/cm to 200 mS/cm. Optional cell constants and low-ion compensation do not remove sample-specific limitations. The current 201T-S listing excludes low-ion water, strong-base samples above pH 12 and highly corrosive liquids. Its probe guide also cautions against frequent testing below pH 2 or above pH 12 and prolonged exposure above 50°C; 0 to 60°C is not a continuous-use guarantee.
The exact manual and specification table state 400 sets, saved and recalled manually, with result, units, temperature and serial number. The product-description bullet conflicts with that figure, so this listing follows the manual and specification table. No timed logging, clock timestamps or electronic export is documented.
No. DO500 is polarographic. It needs polarization, sound membranes, electrolyte and water movement. It automatically compensates temperature and salinity, but pressure compensation is manual. A 30-second 90% response specification at 25°C is not a guarantee of a finished field result in 30 seconds.
Three 50 mL pH buffers and one 50 mL 1413 µS/cm conductivity standard are listed. The DO kits also list three spare DO503 caps, one 30 mL DO502 refill and two cathode-polishing papers. The manual lists two spare AA cells; it does not independently establish an additional installed pair, so four batteries are not promised.
The current US product page and exact model manual specify three years for the instrument and six months for electrodes, both from delivery. Coverage concerns defects attributable to Apera, with repair or replacement at its option. Exclusions include normal wear, improper use, transport/storage damage, failure to follow instructions or preventive maintenance, unauthorized changes or repair, and external causes. No separate extended warranty term for membranes or refill solution is stated; routine consumable replacement is not promised.
Review SX736 AI468 against the sample, supplied electrode and documented method. Return to the live price and ordering controls when the configuration fits your work.
Apera Manufacturer Limited Warranty
Keep the exact product and purchase details when requesting service. Follow the stated model-specific warranty conditions and product instructions; the instrument and supplied sensors or consumables can have different coverage terms.
Use the meter and electrode documents for their stated model and scope. Regional editions are labeled where supplied; software, probe and packing details can differ between revisions. Check the supplied instrument markings when sources describe different hardware revisions. Each card opens a verified manufacturer PDF in a new tab.
email us here: info@bluetigerscientific.com or live chat
Shipping and taxes calculated at checkout. Your quote list is separate.
Review your items, then tell Blue Tiger Scientific where your order is going. Your shopping cart stays unchanged.
International enquiries are welcome. Pricing, shipping availability, freight and any import requirements are confirmed manually.
Prices shown are for reference. We’ll confirm pricing, availability and shipping in your quote.
For eligible Accucold / Summit equipment. White glove and inside delivery are not included and are quoted separately. Shipping availability and final freight are confirmed with your quote, including Alaska, Hawaii and mixed-product requests.
Optional preferences only. Terms must be confirmed by Blue Tiger Scientific; no payment is collected and no terms are approved by this request.
Loading the security check…
* Required. Contact details and eligible quote-cart items will be submitted to Blue Tiger Scientific for this enquiry. No payment is collected.
Which would you like to open?
Review items for online purchase.
Review items and request a quote.
Your two carts stay separate. Nothing is moved or combined.