Give us a call at (833)928-4437
A new conductivity probe for your pocket tester.
Preparing Quote Cart. Your shopping cart stays unchanged.
APERA INSTRUMENTS / EC60-E / AI1209
Apera EC60-E replaces the conductivity probe on EC60 and can be used on PC60 for conductivity measurements. Its brush-resistant platinum-black sensor and built-in temperature sensor support conductivity, with TDS and salinity calculated by the host. It is designed for regular aqueous samples—not for pH measurement.
This page supplies one EC60-E replacement probe (AI1209), not a complete tester. Apera’s probe warranty is six months from delivery.
Only published products are linked. Other cards provide model-comparison information and do not indicate current storefront availability.
Published families link to their product pages. Other models are shown for comparison only; confirm the exact connector and host before choosing.
PH60-E · PH60F-E · PH60S-E
PH60-series pH probes
Standard, flat or spear E probes for PH60, PH60F, PH60S and PC60.
ORP60-E
ORP60-E standard probe
Platinum ORP/temperature probe for ORP60; also listed for PH60, PH60F and PH60S.
EC60-E
EC60-E standard conductivity probe
Platinum-black conductivity/temperature probe for EC60 and PC60; no pH sensor.
PC60-E
PC60-E standard combination probe
pH glass plus platinum-black conductivity and temperature sensing for PC60.
SX615 · SX625
SX610 / SX620 pens
Slim 10 mm glass bulb probes with temperature sensor.
GS1-E · GS2-E · GS3-E · GS4-E · GS5-E
GroStar pen probes
pH, soil spear, EC, pH/EC combo and ORP probes.
EC60-E fits EC60 and PC60. The regional EC60-E sheet also names EC5 and PC5. On PC60 this conductivity-only probe does not retain pH measurement; choose PC60-E for the combined role.
See Alternative format: EC60-DE for the separately sold format documented for EC60-Z and PC60-Z, with model-specific Premium support. DE compatibility does not establish E compatibility with ZenTest bodies.
Apera’s brush-resistant platinum-black sensor is designed for conductivity measurement over the 0–20 mS/cm probe range.
The 30 kΩ thermistor supplies temperature sensing for compatible host operation. Compensation settings remain host-dependent.
TDS and salinity are derived from conductivity by the host tester. A replacement probe does not add unsupported display modes or pH measurement.
EC60-E specifies 0–20 mS/cm, K=1, a platinum-black conductivity sensor, a 30 kΩ thermistor and 0–50°C operation. Use the host’s supported units, temperature compensation and calibration procedure.
See the Product Specifications tab for the exact-model specifications and ordering information.
Rinse away sample residues after use. For EC60-E on EC60, the regional EC60 host manual specifies some distilled or tap water in the cap for storage. This conductivity-only storage routine must not be copied to pH-only or PC60-E combination probes.
If contaminated, the EC60 host manual permits a soft brush and soapy water, followed by rinsing. Do not rub or polish the black pins: this can peel off the platinum-black coating. Keep electrical contacts dry. The EC60-E sheet requires at least two-point calibration after installation.
Use 1413 µS/cm and 12.88 mS/cm conductivity standards as directed by the supported host’s two-point calibration procedure.
Apera’s product page lists EC60 and PC60. The regional exact-probe instruction sheet additionally names EC5 and PC5; this is a specific model list, not a general Series 60 or ZenTest interchangeability claim.
No. EC60-E is a conductivity/temperature probe. Use PC60-E if you need pH and conductivity from the same installed probe.
It uses a brush-resistant platinum-black conductivity sensor with K=1 and a built-in 30 kΩ thermistor.
The exact EC60-E sheet specifies 0–20 mS/cm and 0–50°C operation. The host determines system accuracy, units and compensation settings.
No. TDS/salinity calculations and other display modes belong to the host. Installing EC60-E does not add pH or unsupported functions.
Remove the cap, unscrew the probe ring, unplug the old probe, align and seat the replacement, then tighten the ring. Keep the electrical contacts clean and dry. The EC60-E instruction sheet requires at least two-point calibration after replacement, using the supported host’s standards and procedure.
Rinse after use. The EC60 host manual allows a soft brush and soapy water for contamination, but says not to rub the black pins because the platinum-black coating can peel off.
The regional EC60 host manual specifies some distilled or tap water in the cap for this conductivity-only probe. This is not pH-probe storage advice and must not be applied to the PC60-E combination sensor.
Order EC60-E (AI1209) for the supported tester and sensing role listed above. This is one replacement probe.
APERA INSTRUMENTS, LLC warrants this product to be free from defects in material and workmanship for SIX MONTHS from the delivery.
This limited warranty does NOT cover accidental damage, improper use, normal wear and tear, transportation, storage, failure to follow the product instructions, unauthorized maintenance, modification or repair, or external causes beyond Apera's reasonable control.
Full official PDFs are linked below. Exact-probe instruction sheets cover the named E model; host manuals are labelled separately for operation, calibration and care. Host replacement listings do not broaden E-to-Z compatibility.
email us here: info@bluetigerscientific.com or live chat
Shipping and taxes calculated at checkout. Your quote list is separate.
Review your items, then tell Blue Tiger Scientific where your order is going. Your shopping cart stays unchanged.
International enquiries are welcome. Pricing, shipping availability, freight and any import requirements are confirmed manually.
Prices shown are for reference. We’ll confirm pricing, availability and shipping in your quote.
Optional preferences only. Terms must be confirmed by Blue Tiger Scientific; no payment is collected and no terms are approved by this request.
Loading the security check…
* Required. Contact details and eligible quote-cart items will be submitted to Blue Tiger Scientific for this enquiry. No payment is collected.