Apera PH800 Laboratory Benchtop pH Meter Kit | AI521

$442.00

    One combined probe. A guided calibration. A clearer daily record.

    • 201T-F pH/temperature electrode with BNC and RCA plugs
    • Selectable 0.1 or 0.01 pH display resolution
    • One to three calibration points with slope feedback
    • 500 stored data sets and USB output; software revision matters

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

Preparing Quote Cart. Your shopping cart stays unchanged.

APERA INSTRUMENTS / PH800 / AI521

A practical pH bench, with temperature in the same probe

PH800 pairs a 0.01 pH-resolution benchtop meter with the 201T-F pH/temperature electrode, a flexible holder and starter buffers. It suits routine measurements in compatible general water solutions, with calibration guidance, electrode-slope feedback and timestamped storage to support a repeatable daily method.

Choose PH800 for routine aqueous pH work that benefits from a fixed bench setup, integrated temperature sensing and saved readings. It is a pH/mV instrument, not a conductivity or dissolved-oxygen meter.

Apera PH800 with 201T-F electrode and flexible holder
PH800 with its combined pH/temperature electrode. The beaker is illustrative; see the kit list for included items.
Which 800 or 820 benchtop kit fits my work?Click or tap to compare all six benchtop configurationsClick or tap here to close the guide

Each card opens that model page. Match the supplied electrode and documented sample method. Different configurations are not interchangeable upgrades, and optional accessories are not supplied unless the exact kit contents say so.

Choose the electrode for the sample.

Two connections, one measuring body

The supplied 201T-F contains both the pH sensing system and a temperature sensor. Its blue BNC connector carries pH; the black RCA plug carries temperature. Connecting both enables automatic temperature compensation. This is not the separate LabSen 211 plus MP500 arrangement supplied with PH820.

Calibration that shows its progress

PH800 supports one to three automatic calibration points. USA and NIST buffer series are selectable, and the screen indicates recognized buffers and completed calibration ranges. Recognizing several buffer values does not make this a five-point calibration instrument. Electrode slope provides useful diagnostic context rather than a blanket guarantee of sample accuracy.

Read the status, not only the digits

The backlit display presents pH and temperature together. A stable-reading icon helps identify when to record a result, while HOLD indicates the optional auto-lock mode. M+ and RM distinguish saving from recalling data. Check ATC or MTC before a run so that an unplugged temperature connector does not silently change your method.

A serviceable general-water sensor

The 201T-F uses a polycarbonate body, lithium-glass sensing membrane and non-refillable gel reference. Its protective shield helps guard the bulb but does not make glass unbreakable. The current kit description calls the junction dual ceramic-fiber, while the electrode manual calls it single fiber ceramic; confirm construction if that distinction is critical.

Instrument range is not electrode permission

The current 201T-F probe page specifies 0 to 13 pH and 0 to 60°C, while its linked manual lists 0 to 14 pH and 0 to 80°C. Use the narrower published probe limits until the supplied revision is confirmed. The manual cautions against prolonged testing above 50°C.

Know what this configuration supplies.

Included item Order contents
PH800 benchtop meter One instrument, US kit SKU AI521.
201T-F pH/temperature electrode One combined sensor with BNC/RCA connections and storage bottle.
pH calibration buffers pH 4.00, 7.00 and 10.01; the linked manual lists 50 mL each.
Flexible electrode holder Supports the probe over a sample vessel.
USB cable and PC-Link media Current kit lists a flash drive; legacy manual calls it a disk. Software compatibility must be checked.
Power adapter US input 120 V AC, 60 Hz; output 9 V DC, 300 mA.
Instruction manual and quick guide Setup, calibration and maintenance instructions. Sample beakers and computer shown in imagery are illustrative.
201T-F combined pH and temperature electrode with BNC and RCA plugs
The blue BNC plug carries pH and the black RCA plug carries temperature. One electrode body contains both sensors; the storage bottle keeps the sensing end wet.

Match the sample and preparation method.

Routine aqueous quality checks

A useful station for repeated checks of compatible process-water samples, prepared aqueous solutions and laboratory rinses where a defined sampling method is available. Keep sample identity, temperature and calibration records together when comparing results across shifts.

Teaching and method training

The visible calibration and stability indicators help demonstrate the difference between an apparent reading and a settled measurement. A fixed holder also makes immersion depth and handling easier to repeat without treating the electrode as a stirring rod.

Special samples need a different electrode

The 201T-F manual cautions against extended hot testing above 50°C, frequent strong acid or alkaline exposure, low-ionic-strength water and liquids that attack polycarbonate. Pure water, viscous foods, solvents and solids require a specifically selected sensor and validated method, not simply a different meter setting.

Original 800 Series LCD schematic with numbered status indicators
Original series-manual LCD schematic, not a measurement photograph. 1 parameter; 2 reading; 3 USB connection; 4 timed storage; 5-6 date/time; 7 measurement unit; 8 calibration unit; 9 calibration value or record number; 10 temperature unit; 11 M+ save or RM recall; 12 temperature; 13 ATC automatic or MTC manual compensation; 14 completed calibration; 15 HOLD; 16 stable reading. Conductivity legends belong to other series models, not this pH meter.

Build a repeatable three-buffer routine

Secure the flexible holder on a stable bench, connect the supplied power adapter, and attach both electrode plugs. Loosen the storage-bottle cap before removing the electrode. Inspect the sensing end and shield, rinse with purified water and remove excess water gently. The 201T-F manual calls for a three-point calibration after fitting a new electrode.

Use fresh portions of appropriate buffer in clean vessels, not the stock bottles. Rinse between solutions, remove excess rinse water without rubbing the sensing glass, and allow the electrode and solution to reach a stable temperature. A buffer’s actual pH changes with temperature; select the correct buffer series rather than treating every bottle labelled neutral as interchangeable.

For the standard three-point routine, begin with pH 7.00 and proceed through the appropriate acidic and alkaline buffers, normally 4.00 and 10.01. Wait for recognition and stability before confirming each point. Investigate a calibration error or poor slope rather than repeatedly accepting it. Use buffers that bracket the expected sample range when possible.

Immerse the sensing end and junction adequately in the sample, move gently to release trapped air, then allow the reading to settle. Record temperature with pH. Automatic compensation corrects electrode response to temperature; it does not convert the sample’s chemistry to an assumed pH at another temperature. Keep calibration and sample temperatures reasonably close.

Understand the sensing end and connections.

201T-F sensing end inside the protective shield
Close view of the actual 201T-F sensing end. The shield helps protect the glass but does not make it unbreakable. Keep the glass and junction clean and adequately immersed.

Keep the sensing glass wet and the connectors dry

Rinse before and after each measurement. Do not rub the glass membrane to dry it. Clean deposits with a method suitable for the contaminant and the electrode’s materials, following the exact probe manual. If response becomes slow or calibration cannot be reproduced, clean, condition and recalibrate before deciding whether replacement is needed.

The 201T-F gel reference is not user-refillable. Store its sensing end in Apera 3M KCl storage solution, AI1107, in the storage bottle; replace contaminated solution. Do not use distilled, deionized, RO or tap water for storage. Rinse water is for cleaning between readings, not for long-term soaking.

Keep BNC and RCA connections clean and dry. IP54 describes splash and dust protection, not an immersible instrument. Keep the meter and adapter clear of spills, and turn the meter off before disconnecting power. Glass electrodes age even when maintained correctly; no fixed service life is promised.

Original 201T-F protective sensor shield illustration
The manufacturer’s original shield illustration shows how the guard surrounds the sensing end. Remove residues carefully; do not rub or strike the glass membrane.
Original PH800 LCD stability, calibration and memory guide
Original manufacturer display guide: the smiling icon marks a stable reading; L, M and H mark completed calibration ranges; M+ saves and RM recalls. ATC indicates automatic temperature compensation.

Read the display. Check the measurement mode.

Use the physical controls for measurement, calibration and the documented record functions. Check units, temperature, calibration indicators and the selected parameter before assigning a result to a sample. Screen values in photographs are illustrative, not sample results. The model-specific record and output workflow is explained below.

Use calibration solutions and settings appropriate to the parameter and fitted electrode. Temperature compensation does not make different sample temperatures, preparation methods or chemistry equivalent. Follow the model-specific meter and electrode instructions before extending the method to a new sample.

500 records, with a revision-aware USB workflow

PH800’s current kit page and linked manual agree on 500 stored groups. The manual lists record number, date, time, measurement, unit and temperature, with manual saving and timed storage. Check the internal clock before a batch. Export records before clearing memory and preserve sample identifiers in your own laboratory record.

The linked instrument manual documents a legacy Windows PC-Link program, USB driver and COM-port workflow, with export to an Excel file. The current US kit page also advertises browser-based PC-Link. These are different software generations: confirm support for the delivered meter revision, operating system, browser and required export fields before committing to a computer workflow. No software installation or runtime compatibility test is implied here.

The current page additionally describes calibration date, slope, offset and serial-number reporting. Treat these as revision-dependent software fields, not as proof that every older unit stores them in every record. In the legacy manual’s connected real-time mode, readings go to the PC rather than remaining in meter memory. Plan record retention accordingly.

PH800 USB connection and legacy PC-Link example
The USB icon indicates a connected computer. This original PH800 graphic shows legacy Windows PC-Link and 500-record storage, not a tested current-browser workflow. Computer not included. The manufacturer’s “GLP Data Management” wording describes recording functions, not GLP compliance, a validated audit trail or validated LIMS integration.

Compare a TestBench benchtop alternative.

Alternative format: Apera PH950 Benchtop pH Meter Kit with LabSen 211 and 606 TestBench. The linked TestBench model has a different bench arrangement, record workflow and kit specification. Compare the exact supplied electrode, accessories and data features rather than assuming similar model names mean identical hardware or software.

PH800 buyer questions

Is the temperature probe separate?

No. The supplied 201T-F combines pH and temperature in one body, with separate BNC and RCA plugs. A separate temperature probe may be needed if you later choose a pH-only replacement electrode.

Does PH800 measure conductivity?

No. PH800 measures pH, mV and temperature. The shared series manual also covers conductivity models, but their EC, TDS, salinity and resistivity functions do not belong to PH800.

Can I calibrate at five points?

No. PH800 supports one to three automatic calibration points. The list of recognized USA buffer values describes buffer recognition, not the number of points stored in one calibration.

Can the kit test at 100°C?

The current 201T-F probe page specifies 0 to 13 pH and 0 to 60°C, while its linked manual lists 0 to 14 pH and 0 to 80°C. Use the narrower published probe limits until the supplied revision is confirmed. The manual cautions against prolonged testing above 50°C. The instrument display and compensation limits remain separate specifications.

Does USB guarantee compatibility with my computer?

No. The manual describes legacy Windows PC-Link; the current page describes a browser-based option. Confirm meter revision and current host requirements. Computer hardware is not included.

Is an ORP electrode included?

No. The mV mode can be used for ORP with an appropriate separately purchased electrode. The included 201T-F is a pH/temperature electrode, not an ORP sensor.

Is the reference electrolyte refillable?

No. The 201T-F has a gel reference. Maintain external storage solution in its bottle, but do not attempt to refill the electrode body.

What warranty applies?

Apera’s current US instrument manual provides a three-year limited warranty for the instrument and six months for the probe, measured from delivery, for defects attributable to Apera. Repair or replacement is at Apera’s option. Exclusions include accidental damage, transportation, storage, improper use, failure to follow instructions or preventive maintenance, unauthorized repair and normal wear. See the linked manual for the complete terms.

Confirm the exact configuration before ordering.

Review PH800 AI521 against the sample, supplied electrode and documented method. Return to the live price and ordering controls when the configuration fits your work.

Back to price and ordering ↑
    SpecificationDetail
    Model / SKUPH800 / AI521
    ParameterspH, mV, temperature
    Meter pH range-2.00 to 19.99 pH
    pH resolution0.1 / 0.01 pH
    Instrument pH accuracy±0.01 pH ±1 digit
    Automatic pH calibration1 to 3 points; USA / NIST series
    Supplied electrode201T-F; BNC/RCA; integrated temperature sensor
    Electrode rangeThe current 201T-F probe page specifies 0 to 13 pH and 0 to 60°C, while its linked manual lists 0 to 14 pH and 0 to 80°C. Use the narrower published probe limits until the supplied revision is confirmed. The manual cautions against prolonged testing above 50°C.
    pH temperature compensation0 to 100°C, automatic or manual; probe limits still apply
    Temperature display / accuracy0.1°C; manual specifies ±0.5°C ±1 digit
    Temperature range qualificationSupplied 201T-F: use 0 to 60°C following the narrower current probe page until the delivered revision is confirmed. Its manual lists 0 to 80°C and cautions against prolonged use above 50°C. Meter display and compensation limits are separate.
    mV range / resolution / accuracy±1999 mV / 1 mV / ±0.1% FS ±1 digit
    Memory / output500 groups / USB
    ProtectionIP54, splash and dust protection; not immersion
    Power120 V AC 60 Hz input; 9 V DC 300 mA output
    Limited warrantyInstrument: 3 years; probe: 6 months from delivery; exclusions apply

    Ordering Information

    Ordering detailConfiguration
    Manufacturer catalogAI521
    Blue Tiger SKUBTS-AI521
    Order unit1 exact 800 or 820 Series kit
    Order contentsPH800 benchtop meter: One instrument, US kit SKU AI521.; 201T-F pH/temperature electrode: One combined sensor with BNC/RCA connections and storage bottle.; pH calibration buffers: pH 4.00, 7.00 and 10.01; the linked manual lists 50 mL each.; Flexible electrode holder: Supports the probe over a sample vessel.; USB cable and PC-Link media: Current kit lists a flash drive; legacy manual calls it a disk. Software compatibility must be checked.; Power adapter: US input 120 V AC, 60 Hz; output 9 V DC, 300 mA.; Instruction manual and quick guide: Setup, calibration and maintenance instructions. Sample beakers and computer shown in imagery are illustrative.

    Apera Manufacturer Limited Warranty

    Apera states a limited three-year warranty for the instrument and a six-month warranty for the supplied probe from delivery, for defects attributable to the manufacturer and subject to its stated conditions.

    Normal wear, accidental damage, improper use, transport or storage damage, failure to follow instructions, unauthorized repairs or modifications and other external causes are excluded. Keep the exact product and purchase details when requesting service.

    PH800 official documents

    Use the meter and electrode documents for their stated model and scope. Regional editions are labeled where supplied; software, probe and packing details can differ between revisions. Check the supplied instrument markings when sources describe different hardware revisions. Each card opens a verified manufacturer PDF in a new tab.

Questions about a product? Ask us here.