Skip to content

Global shipping available. Please request a quote to receive customized pricing for your region.

Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes LabSen 241-180 – NMR tubes and slim containers pH Electrode Apera Instruments
Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes LabSen 241-3 – micro-volume samples pH Electrode Apera Instruments
Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes LabSen 241-3SP – protein-containing micro samples pH Electrode Apera Instruments
Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes pH Electrode Apera Instruments
Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes LabSen 241-180 – NMR tubes and slim containers pH Electrode Apera Instruments
Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes LabSen 241-3 – micro-volume samples pH Electrode Apera Instruments
Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes LabSen 241-3SP – protein-containing micro samples pH Electrode Apera Instruments
Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes pH Electrode Apera Instruments

Apera LabSen 241-3 / 241-3SP / 241-180 Refillable Micro pH Electrode for Micro-Volume Samples and NMR Tubes

$229.00

    Micro-volume pH: 30 µL for 241-3/241-3SP; 50 µL for 241-180.

    • Microplates, tubes, NMR
    • pH 0–13, 0–100°C
    • Versions: 241-3 BNC; 241-3SP BNC; 241-180 BNC
    • Ø3 mm cylindrical tips; 241-3SP uses Protelyte for refill and storage

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

APERA INSTRUMENTS / LABSEN 241-3 · 241-3SP · 241-180 / AI3104 / AI3151 / AI3161

Micro-volume pH.
30–50 µL by model.

LabSen 241-3, 241-3SP and 241-180 are refillable micro pH electrodes for samples too small for a standard probe. The 241-3 has a Ø3 mm tip on a Ø4 mm stainless steel protective shaft for microplates, microcentrifuge tubes and microbiological samples from 30 µL. The 241-3SP is the same design with Protelyte electrolyte for protein-containing samples such as serum. The 241-180 has a Ø3 × 180 mm measuring tip and a Ø12 × 60 mm body for NMR tubes and other slim, deep containers, from 50 µL. All three use a BNC connector.

It reads pH 0 to 13 from 0 to 100°C. Choose the version for your sample from the menu above. Apera warrants LabSen electrodes for six months from delivery.

Apera LabSen 241-3 · 241-3SP · 241-180 pH electrode
Apera LabSen 241-3 · 241-3SP · 241-180.
Choose your LabSen electrode: small samples and solidsClick or tap to compare 6 electrode familiesCompare the models below - click or tap here to close

Only published products are linked. Other cards provide model-comparison information and do not indicate current storefront availability.

Pick the version
for your sample.

Version Connector and build
LabSen 241-3 · AI3104 BNC connector, no temperature sensor; refillable; ceramic junction, 3M KCl electrolyte.
LabSen 241-3SP · AI3151 BNC connector, no temperature sensor; refillable; ceramic junction, Protelyte electrolyte.
LabSen 241-180 · AI3161 BNC connector, no temperature sensor; refillable; ceramic junction, 3M KCl electrolyte.

BNC versions plug into any pH meter with a BNC input. They have no temperature sensor, so pair them with a separate temperature probe such as Apera’s MP500 where compatible with the meter, for automatic temperature compensation; manual temperature entry is an option only if the meter supports it at the sample temperature. Electrode ratings do not extend the meter’s or separate temperature probe’s operating or compensation limits.

What makes it work.

Micro cylindrical membrane

Low-resistance LabSen glass at a Ø3 mm tip.

Protective shaft (241-3, 241-3SP)

A Ø4 mm stainless steel shaft protects the slim glass tube. Apera states volumes as low as 15 µL with its micro container.

Long-Life reference system

Ceramic junction and refillable reference: 3M KCl, or Protelyte on the 241-3SP.

Check the specifications.

All versions share: pH range 0–13; temperature range 0–100°C; glass membrane micro, cylindrical; junction ceramic.

Reference electrolyte: 241-3 3M KCl; 241-3SP Protelyte; 241-180 3M KCl.

Dimensions: 241-3 and 241-3SP have a Ø3 × 16 mm exposed measuring tip and a Ø4 × 54 mm stainless-steel protected section; the manual lists the body as Ø12 × 130 mm. The 241-180 has a Ø3 × 180 mm measuring tip and a Ø12 × 60 mm body: 180 mm is the tip length, not the entire electrode.

The full specification table is in the Product Specifications tab.

Apera LabSen 241-3 dimension drawing
LabSen 241-3 simplified manufacturer illustration. Its 70 mm measuring end includes the manual’s 16 mm section at 3 mm diameter and 54 mm protective section at 4 mm diameter. Do not infer a uniform 3 mm diameter along the full 70 mm; use the detailed specification and supplied-revision instructions for vessel fit.

Store it wet.
Keep it clean.

Keep the tip wet in storage solution in its cap between uses; the 241-3 and 241-180 use 3M KCl storage solution.

Never store a pH electrode in pure, deionized or distilled water.

The 241-3, 241-3SP and 241-180 are refillable: keep the fill hole open while measuring and top up the reference electrolyte when the level drops.

Use 3M KCl to refill and store the 241-3 and 241-180. The 241-3SP requires Protelyte for both refill and storage; after cleaning, soak its tip in Protelyte for at least 1 hour, then rinse and recalibrate. Do not apply the other versions’ KCl instructions to the 241-3SP.

Rinse with distilled or deionized water between samples. For stubborn deposits, use Apera electrode cleaning solution, then recondition and recalibrate.

Related items.

Apera ships this electrode in these kits: PH700-MS (LabSen 241-3SP).

Calibrate with fresh pH 7.00, pH 4.00 and pH 10.01 buffers.

LabSen 241-3 · 241-3SP · 241-180 buyer questions

What is the LabSen 241-3 / 241-3SP / 241-180 made for?

LabSen 241-3, 241-3SP and 241-180 are refillable micro pH electrodes for samples too small for a standard probe.

Which version do I need?

All use BNC without a temperature sensor. Choose 241-3 (AI3104) for general micro samples, 241-3SP (AI3151) for protein-containing samples such as serum, or 241-180 (AI3161) for NMR tubes and other slim, deep vessels.

Is it refillable?

Yes. Refill LabSen 241-3 and 241-180 with 3M KCl, but refill 241-3SP with Protelyte. Keep the fill hole open during measurement.

How should I store it?

Store LabSen 241-3SP in Protelyte; after cleaning, soak it in Protelyte for at least 1 hour, then rinse and recalibrate. Store 241-3 and 241-180 in 3M KCl. Keep the tip wet; never store any version in pure, deionized or distilled water.

How little sample can I measure?

LabSen 241-3 and 241-3SP are for samples from 30 µL; 15 µL requires Apera’s micro container. LabSen 241-180 requires at least 50 µL. Vessel shape must allow the sensing end and junction to contact the sample.

Will the tip fit my microtube or NMR tube?

For 241-3/241-3SP, allow for a Ø3 × 16 mm exposed tip followed by a Ø4 × 54 mm stainless-steel protected section, not a uniform Ø3 mm shaft. The 241-180 has a Ø3 × 180 mm measuring tip and a Ø12 × 60 mm body; 180 mm is not its total length.

Do these micro electrodes provide automatic temperature compensation?

No built-in temperature sensor is fitted. Use a meter-compatible separate sensor, or manual temperature entry where supported at the actual sample temperature. The offered electrode limits remain pH 0–13 and 0–100°C; these do not expand the host’s limits.

What warranty applies?

Apera warrants LabSen electrodes against defects in material and workmanship for six months from delivery. Accidental damage, improper use and normal wear are excluded.

Confirm the exact configuration before ordering.

Check the version, connector and catalog number (AI3104 / AI3151 / AI3161) against your meter and samples, then return to the live price and ordering controls.

Back to price and ordering ↑
    SpecificationDetail
    ModelLabSen 241-3 / 241-3SP / 241-180
    ConnectorBNC
    Temperature sensor241-3: none; 241-3SP: none; 241-180: none
    pH range0–13
    Temperature range0–100°C
    Body241-3: Lead-free glass, stainless steel protective shaft; 241-3SP: Lead-free glass, stainless steel protective shaft; 241-180: Lead-free glass
    Glass membraneMicro, cylindrical
    JunctionCeramic
    Reference systemLong-Life
    Reference electrolyte241-3 / 241-180: 3M KCl for refill and storage. 241-3SP: Protelyte for refill, storage and reconditioning after cleaning; do not substitute the other versions’ KCl care instructions.
    RefillableYes
    Dimensions241-3 / 241-3SP: Ø3 × 16 mm exposed measuring tip; Ø4 × 54 mm stainless-steel protected section; body Ø12 × 130 mm. 241-180: Ø3 × 180 mm measuring tip; Ø12 × 60 mm body. The 180 mm value is not entire-electrode length.
    Membrane impedance<450 MΩ
    Cable3 ft
    Manufacturer warrantySix months from delivery against defects in material and workmanship

    Ordering Information

    Ordering detailConfiguration
    Manufacturer catalogAI3104 / AI3151 / AI3161
    Blue Tiger SKUBTS-AI3104 / BTS-AI3151 / BTS-AI3161
    VersionsLabSen 241-3 – micro-volume samples: AI3104; LabSen 241-3SP – protein-containing micro samples: AI3151; LabSen 241-180 – NMR tubes and slim containers: AI3161
    Order unit1 electrode

    Warranty Information

    APERA INSTRUMENTS, LLC warrants LabSen electrodes to be free from defects in material and workmanship for SIX MONTHS from the delivery.

    This limited warranty does NOT cover accidental damage, improper use, normal wear and tear, transportation, storage, failure to follow the product instructions, unauthorized maintenance, modification or repair, or external causes beyond Apera's reasonable control.

    LabSen 241-3 · 241-3SP · 241-180 official documents

    Official Apera documents are labeled by model and scope below. Each whole card opens a PDF in a new tab.

    Document scope and limits. Each manual applies to its named model. The 241-3SP specifically requires Protelyte for refill, storage and reconditioning; 241-3/241-180 use 3M KCl. These manuals state pH 0–14, while current US product pages specify 0–13; this offer retains the narrower 0–13 range.

Questions about a product? Ask us here.

email us here: info@bluetigerscientific.com or live chat