Choose infrared by method. Keep the results in context. TN480 AI486 for the defined measurement workflow.
Preparing Quote Cart. Your shopping cart stays unchanged.
T200-1 is the 20/100/400/800 NTU set documented for TN400 and TN480.
Swipe or use the arrows to browse.
APERA INSTRUMENTS / TN480 / AI486
TN480 combines an infrared LED optical method with selectable TruRead and a Windows reporting workflow. Its 860 ± 30 nm source is the central distinction from the white-light TN420 and TN500, not a cosmetic change of model number.
Choose TN480 when your required procedure calls for the manufacturer-stated ISO 7027 / DIN EN 27027 infrared method. The source is also positioned for appropriate colored samples, but no optical design eliminates every matrix interference.
0-1000 NTU
Instrument range
Method/reporting scope is a separate requirement.
860 ± 30 nm
Infrared LED
Match the required optical method.
Windows records
PC-Link + selectable TruRead
Capacity depends on mode.
Start with the required optical method, then the recording workflow. Each card opens that product page. T500-1 is replacement standards only, not a fourth meter.
AI482
White light / essential workflow
Fixed 10-reading average. USB charging; no documented record export.
AI483
White light / recorded results
Selectable TruRead and Windows PC-Link data export.
AI481
Infrared / essential workflow
ISO 7027 optical method. Four AA batteries; readings recorded by hand.
AI486
Infrared / recorded results
ISO 7027 optical method. Four AA batteries and Windows PC-Link.
AI4205
STANDARDS ONLY / TN420 + TN500
Four sealed nonzero AMCO vials. No meter or zero. TN480 specifies T200-1.
OPTICAL METHOD BEFORE UNIT LABEL
Infrared LED, 860 ± 30 nm, with manufacturer-stated ISO 7027 / DIN EN 27027 positioning.
Tungsten white light, 400-600 nm, with manufacturer-stated EPA 180.1 positioning.
Optics, geometry, sample matrix and the required method govern interpretation. A unit label is not a universal conversion rule.
Apera positions this infrared design for appropriate colored samples. It is not immune to every interference, and it does not become an EPA white-light meter because a calibration material carries an EPA-related claim. The exact manual describes NTU/FTU display settings; a separately selectable FNU mode is not asserted.
WHY CHOOSE THIS CONFIGURATION
860 ± 30 nm infrared LED and 90° scattering distinguish this setup from a tungsten white-light meter. Match the instrument to the required method before comparing records features.
Selectable TruRead provides individual values, average, minimum and maximum. Changing samples still need consistent mixing, bubble control and timing.
Windows PC-Link and USB support reporting. The power specification is four AA alkaline batteries, not a rechargeable lithium pack or an implied USB-charging feature.
TN480 offers Normal measurement and selectable 5/10/15/20-reading TruRead with average, minimum, maximum and individual values. Its manual gives 10 seconds for a Normal value; the TruRead example uses 20-second intervals.
USB connects to the supplied Windows PC-Link workflow for transfer and Excel-document export. Specific current Windows versions, Mac/mobile support and cloud backup are not established; check your IT/driver requirements before deployment.
Plan storage around the selected sequence, not the largest headline number. The exact manuals give these onboard capacities:
Mixed use changes the available capacity. Export needed records before clearing the instrument; its clear-all function removes stored records rather than offering a selective archival system.
PC-Link supports a separate live-PC workflow. When that workflow is running, the manuals describe measurements going to the computer rather than being stored in the instrument. This is not cloud backup or a verified electronic-signature/audit-trail system.
READ THE TWO NUMBERS DIFFERENTLY
0.01 NTU
Low-range display increment
Other ranges use 0.1 and 1 NTU increments.
2% + stray light
Published reading-error specification
Accuracy: ≤±2% of reading plus stray light; stray light ≤0.02 NTU.
Worked specification example: at 1.00 NTU, the 2%-of-reading term alone is 0.02 NTU before the stray-light term. That arithmetic is not a complete method-uncertainty calculation or a promise of ±0.01 NTU measurement accuracy. Vials, sample preparation, calibration and the matrix remain part of the result.
TN480 meter; five standards at 0/20/100/400/800 NTU; six numbered sample vials; 10 mL silicone oil; microfiber cloth; USB cable; PC-Link software flash drive; four AA alkaline batteries; battery-cover screwdriver; manual and case.
The complete meter kit includes zero. Replacement nonzero standards are a separate purchase; do not read a replacement-set name as all five calibration solutions.
SAMPLE PREPARATION IS PART OF THE MEASUREMENT
Use the specified 25 × 60 mm numbered sample vial and approximately 18 mL of sample. Keep the matching cap paired with it and follow the exact rinsing/mixing procedure.
Fine bubbles can raise readings; settling and light-absorbing material can change them. Keep preparation and waiting times consistent rather than averaging away a method problem.
Keep the vial clean and dry, avoid fingerprints/scratches, align its mark and close the cover. Use oil only as instructed and remove excess.
Work on a flat, level surface, not with the meter in your hand. Never pour sample into or wash the optical holder. The IP67 rating is not a direct-immersion measurement method. Turbidity is not a direct TSS mass measurement or a complete water-safety test.
The kit is supplied with five calibration values and is factory calibrated across its range. The manual permits later calibration with two or more appropriate points bracketing the expected range; local method/QC requirements still govern. Select the correct value because the instrument does not identify a standard automatically.
For low-range work, follow the manual’s zero check, warm-up and numbered-vial guidance. A display increment is not proof that an unchecked zero or contaminated vial is adequate.
The four nonzero AMCO standard vials must stay sealed. Zero-vial opening/refill instructions are different. Do not vigorously shake standards or infer that a generic shelf-life claim permits opening them.
The exact TN480 replacement set is T200-1 for the four nonzero values. T500-1 is assigned to TN420/TN500 and is not represented as compatible with this instrument. Zero refill is a separate item.
Alternative optical method: TN500 white-light recording meter and Alternative optical method: TN420 white-light meter are different method choices, not equivalent replacements for an infrared requirement.
The official family tutorial introduces the workflow. Use the exact TN480 manual for its controls, optical method, standards and software. The video is not a model-specific certification or a replacement for sealed-vial instructions.
TN480 AI486 uses infrared LED optics and is positioned by Apera for ISO 7027 / DIN EN 27027. TN500 uses tungsten white light. Choose the required method first; sharing a turbidity range does not make the methods interchangeable.
No. Instrument optics, geometry and reporting method matter. Source literature uses varying unit labels; the exact manual describes NTU/FTU display settings, not a separately verified FNU selector. Do not convert or substitute methods by changing a label alone.
It holds 200 Normal sets, or 100/61/44/34 TruRead sets at 5/10/15/20 readings respectively. Mixed-mode capacity depends on the records saved. Clear-all is irreversible; export needed records first.
Do not select them for TN480. The manufacturer assigns T200-1 to TN480/TN400 infrared models and T500-1 to TN500/TN420 white-light models. Matching 20/100/400/800 labels alone is not compatibility evidence.
The meter uses four AA alkaline batteries. Its documented USB/PC-Link reporting workflow is Windows-based with software supplied on a flash drive. No built-in rechargeable-lithium specification, Mac/mobile support or modern Windows-version guarantee is established.
The kit lists five standards at 0/20/100/400/800 NTU, six sample vials, 10 mL silicone oil, microfiber cloth, USB cable, software flash drive, four AA batteries, screwdriver, manual and case. A Windows computer is not included.
No. It helps show changing readings and their spread. Use a level surface, clean indexed vials, controlled mixing/timing and bubble handling. Never pour sample into or wash the optical vial holder.
No. Published accuracy is ≤±2% of reading plus stray light, with stray light ≤0.02 NTU. The display increments are 0.01/0.1/1 NTU across ranges. These are not interchangeable specifications or a total method-uncertainty guarantee.
Review TN480 AI486 and return to the live price and ordering controls. Ask Blue Tiger Scientific to clarify the exact configuration, method or documentation requirements before ordering.
APERA INSTRUMENTS, LLC warrants this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at the option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to the responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS for the instrument from the delivery.
This limited warranty does NOT cover any issues due to:
The exact-model manual controls operation and configuration. The family brochure is a supplemental selection reference; it also illustrates other models and accessories. A four-value replacement set is not the complete five-standard meter-kit package. Each card opens a PDF in a new tab.
email us here: info@bluetigerscientific.com or live chat
Shipping and taxes calculated at checkout. Your quote list is separate.
Review your items, then tell Blue Tiger Scientific where your order is going. Your shopping cart stays unchanged.
International enquiries are welcome. Pricing, shipping availability, freight and any import requirements are confirmed manually.
Prices shown are for reference. We’ll confirm pricing, availability and shipping in your quote.
For eligible Accucold / Summit equipment. White glove and inside delivery are not included and are quoted separately. Shipping availability and final freight are confirmed with your quote, including Alaska, Hawaii and mixed-product requests.
Optional preferences only. Terms must be confirmed by Blue Tiger Scientific; no payment is collected and no terms are approved by this request.
Loading the security check…
* Required. Contact details and eligible quote-cart items will be submitted to Blue Tiger Scientific for this enquiry. No payment is collected.
Which would you like to open?
Review items for online purchase.
Review items and request a quote.
Your two carts stay separate. Nothing is moved or combined.