A practical pH station for your bench. PH700 pairs a supplied 201T-F electrode and holder with straightforward calibration and local manual recall.
Preparing Quote Cart. Your shopping cart stays unchanged.
Replace the PH700’s supplied pH/temperature electrode with the 201T-F replacement.
When routine supplies are due for replenishment, use the AI1170 kit for pH 4/7/10 buffers, 3M KCl, cleaning solution, brush and CalPod.
Swipe or use the arrows to browse.
APERA INSTRUMENTS / PH700 / AI501
Build a practical general-water pH station around the PH700 benchtop meter, its supplied 201T-F pH/temperature electrode and simple side-mounted holder. The adapter-powered format keeps the probe, controls and calibration routine together at the bench.
Choose AI501 when stationary testing and local manual save/recall fit the work. It is not the portable PH400S recording/export kit, and it does not include every LabSen specialty electrode pictured in the broader PH700 family.
±0.01 pH
Meter accuracy
Stated accuracy includes ±1 digit; sample-method effects are separate.
1-3 points
Automatic pH calibration
Five supported buffer values do not mean five calibration points.
50 sets
Local manual save/recall
Shared across pH/mV; no documented USB or PC export.
Choose by sample and working format. Each card opens that product page. These are alternative workflows, not interchangeable probes or extra items in the PH700 box.
AI501
General-water bench workflow
201T-F BNC/RCA probe, supplied stand, 50 local manual records.
AI421
Alternative format / portable records
Portable recording and USB workflow. Different housing and electrode interface.
AI311
Alternative format / pocket checks
Compact standalone pH testing, not the PH700 bench configuration.
AI3723
Related workflow / pure water
Dedicated low-ionic-water configuration; not a PH700 replacement electrode.
BENCH CONVENIENCE, NOT A DIFFERENT ACCURACY PROMISE
DC9V adapter power and the supplied holder suit a regular bench position. IP54 splash/dust resistance is not the waterproof field rating of a portable kit.
The 201T-F connects blue BNC to pH/mV and black RCA to temperature. A connector that fits physically is not proof of the correct temperature-sensor match.
Save a stable reading with M+ and recall it with RM. The 50-set memory is a local manual record, not automatic interval logging or a verified GLP audit-trail system.
Apera identifies a Texas Instruments microprocessor in PH700’s signal-processing design. This manufacturer feature illustration adds context to the electronics behind the controls.
The published pH specification, calibration and sample/electrode conditions still define the measurement. A chip brand does not add USB export, automatic interval logging or a certified audit trail to this model.
The manual describes a metal stand base, support bar and electrode clip that can be mounted on either side of the meter. Use the simple supplied arrangement rather than assuming an articulated laboratory arm is part of AI501.
Keep the glass bulb and reference contact appropriately immersed in a representative sample. A beaker used for demonstrations is not a supplied kit component.
EXACT BASE KIT
PH700 plus the 201T-F. The electrode is the base general-water configuration, not a LabSen specialty sensor.
One 50 mL bottle each of pH 4.00, 7.00 and 10.01. These are the supplied standards, not five bottles because the meter recognizes five possible values.
One simple holder, power adapter and instructions. No ORP probe, USB cable, beaker, large articulated arm or carrying case is promised.
The 201T-F has a storage vial/cap. That is not evidence that a separate retail bottle of storage solution is included. Optional related products and alternative PH700 kits are separate purchases.
The included 201T-F is a non-refillable gel-reference electrode for appropriate general-water measurements. The PH700 manual recommends routine samples at pH 2-12 and 10-50°C, with medium ionic strength such as conductivity 100-10,000 µS/cm or salinity below 0.5%.
Those conductivity/salinity figures describe an electrode application context. PH700 does not measure EC or salinity. Low-ionic water, heavily contaminated liquids, strong chemistry and hot-end samples need a separately evaluated electrode and method.
Instrument display and compensation intervals do not override the electrode or sample-method limits. Manufacturer documents have differing temperature endpoints, so this page does not convert the largest meter number into an unrestricted whole-kit sample rating.
MAKE THE TWO LAYERS EXPLICIT
The manual gives automatic/manual compensation over 0-100°C. This is an instrument function, not permission to expose the included electrode to every temperature in that interval.
The manual’s recommended general-water envelope is pH 2-12 and 10-50°C. Select the actual matrix, temperature and method within appropriate probe guidance.
Product-page/manual and standalone-probe endpoints differ. Confirm the exact electrode revision and working conditions before relying on an edge-of-range rating.
Choose the USA or NIST buffer series in the meter settings. The supported USA values are 1.68, 4.00, 7.00, 10.01 and 12.45; the NIST series includes 1.68, 4.01, 6.86, 9.18 and 12.45. Calibration uses at most three points, not five.
Start with the neutral point as instructed: 7.00 for USA or 6.86 for NIST. Add appropriate acidic/alkaline points for the expected sample. After two- or three-point calibration the meter shows slope information, helping the operator review the calibration rather than simply accepting a displayed number.
Use fresh uncontaminated standards and the current manual’s rinsing, stability and calibration steps. Automatic standard recognition does not remove sample preparation, electrode care or the required laboratory quality-control method.
The backlit LCD shows pH and temperature with stability and calibration indicators. The max/min function tracks extrema during its selected measurement period; it is not a claim of a permanent exported history.
M+ manually saves a stable result and RM recalls it. The 50 sets are shared by pH and mV. Stored fields include the record number, measurement, unit, temperature and compensation status; a date/time field is not promised.
At capacity, FUL appears and the manual directs a clear through P4 before further storage. Plan your recording procedure before clearing data. No USB/PC export or automatic interval-recording function is established for PH700.
ORP IS A DIFFERENT SENSOR TASK
With the supplied pH electrode connected, mV mode displays that electrode’s original pH millivolt signal. An ORP measurement requires a separately purchased appropriate ORP electrode, such as the manufacturer-referenced 301Pt-C.
The instrument’s mV range is ±1999 mV with 1 mV resolution. It does not establish that an ORP electrode is included, that any BNC accessory is a validated complete replacement, or that temperature compensation for pH is an ORP compensation claim.
The supplied 201T-F uses blue BNC for pH/mV and black RCA for its 30KΩ temperature sensor. PH700’s IP54 protection is splash/dust resistance, not an immersion or waterproof field rating.
A compatible-looking BNC plug is only one part of selection. Check the electrode chemistry, temperature sensor or manual-compensation arrangement, and the intended sample before choosing an optional probe. The manufacturer’s illustrated alternatives are not extra sensors in this box.
Rinse appropriately between standards and samples, avoid contaminating buffers, and protect the glass tip from abrasion or impacts. Let the reading stabilize under the actual method instead of forcing a result by timing alone.
The 201T-F uses a gel reference and is not a user-refillable reservoir. Keep the tip in its tightly closed storage vial with the specified 3M KCl solution. Tap, RO, distilled and DI water are rinsing waters, not storage substitutes.
Use the current manufacturer instructions for contaminant-specific cleaning, conditioning and recalibration. A dirty, dry or poorly conditioned electrode should not be diagnosed solely by the meter’s last displayed decimal place.
Apera’s PH700 tutorial covers the bench setup, connector arrangement, calibration, measurement, local memory and optional ORP work. Use the current manual for the exact configuration and operating limits.
Alternative format: PH400S portable recording/USB workflow is a different working format and electrode interface, not the PH700 kit with a different label.
Alternative format: PH60 pocket pH tester suits compact standalone checks when a permanent bench station is unnecessary.
Related workflow: PH60-PW specialized pure-water kit addresses a low-ionic-water method. Its external probe is not represented as a mechanically compatible PH700 replacement.
The base kit includes PH700, one 201T-F pH/temperature electrode, one holder, adapter, three 50 mL buffers at pH 4.00/7.00/10.01, manual and quick-start guide. It does not promise a LabSen or ORP probe, USB cable, beaker, large arm or carrying case.
PH700 AI501 is an adapter-powered bench station with a 201T-F BNC/RCA electrode and 50 local manually saved sets. PH400S is a portable recording/export workflow with a different electrode connection. Choose by working format, not a claim that PH700 has superior published pH accuracy.
No. PH700 supports 1-3 calibration points. It can recognize five values in the selected USA or NIST series, but that list is not the number of simultaneous calibration points. Start neutral as directed and select additional points for the sample.
No USB/PC export function is documented. The 50 sets are shared by pH/mV and manually saved/recalled. No automatic interval logging, date/time record field or certified audit-trail system is promised.
No. The 201T-F in mV mode shows its original pH electrode signal. ORP requires a separately purchased appropriate ORP electrode. A ±1999 mV instrument range is not evidence of an included ORP sensor.
Do not assume complete compatibility from a connector alone. The supplied 201T-F uses BNC for pH/mV and RCA for its 30KΩ temperature sensor. Match the electrode, temperature interface and method using manufacturer evidence; the 8-pin portable configuration is different.
The included 201T-F is the general-water configuration. The manual recommends pH 2-12 and 10-50°C with medium ionic strength. Specialty matrices and extreme conditions need an appropriate separately evaluated electrode/method. Meter compensation range does not override the supplied probe.
PH700 is IP54 splash/dust-resistant, not an IP57 waterproof field kit. Keep the meter body out of samples and use the electrode correctly. The supplied simple holder is a bench convenience, not permission to immerse the instrument.
Review the PH700 AI501 base configuration and return to the price and ordering controls. Blue Tiger Scientific can help clarify the exact electrode and method requirements before ordering.
APERA INSTRUMENTS, LLC warrants this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at the option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to the responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS for the instrument and SIX MONTHS for the probe from the delivery.
This limited warranty does NOT cover any issues due to:
The current manufacturer-linked manual covers the base kit, connectors, calibration, local records, electrode care, optional ORP and warranty. The whole card opens the PDF in a new tab.
email us here: info@bluetigerscientific.com or live chat
Shipping and taxes calculated at checkout. Your quote list is separate.
Review your items, then tell Blue Tiger Scientific where your order is going. Your shopping cart stays unchanged.
International enquiries are welcome. Pricing, shipping availability, freight and any import requirements are confirmed manually.
Prices shown are for reference. We’ll confirm pricing, availability and shipping in your quote.
For eligible Accucold / Summit equipment. White glove and inside delivery are not included and are quoted separately. Shipping availability and final freight are confirmed with your quote, including Alaska, Hawaii and mixed-product requests.
Optional preferences only. Terms must be confirmed by Blue Tiger Scientific; no payment is collected and no terms are approved by this request.
Loading the security check…
* Required. Contact details and eligible quote-cart items will be submitted to Blue Tiger Scientific for this enquiry. No payment is collected.
Which would you like to open?
Review items for online purchase.
Review items and request a quote.
Your two carts stay separate. Nothing is moved or combined.