Apera TN480 Portable Infrared Turbidity Meter with Data Logging Portable Turbidity Meter Apera Instruments
Apera TN480 Portable Infrared Turbidity Meter with Data Logging Portable Turbidity Meter Apera Instruments
Apera TN480 Portable Infrared Turbidity Meter with Data Logging Portable Turbidity Meter Apera Instruments
Apera TN480 Portable Infrared Turbidity Meter with Data Logging Portable Turbidity Meter Apera Instruments
Apera TN480 Portable Infrared Turbidity Meter with Data Logging Portable Turbidity Meter Apera Instruments
Apera TN480 Portable Infrared Turbidity Meter with Data Logging Portable Turbidity Meter Apera Instruments

Apera TN480 Portable Infrared Turbidity Meter with Data Logging

$1,050.00

    Choose infrared by method. Keep the results in context. TN480 AI486 for the defined measurement workflow.

    • Infrared LED, 860 ±30 nm; 90° nephelometric scattering
    • Normal, or selectable 5/10/15/20-reading TruRead with average, minimum, maximum and individual values
    • Windows PC-Link; mode-dependent onboard capacity
    • Includes five standards and six sample vials

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

Preparing Quote Cart. Your shopping cart stays unchanged.

APERA INSTRUMENTS / TN480 / AI486

Choose infrared by method.
Keep the results in context.

TN480 combines an infrared LED optical method with selectable TruRead and a Windows reporting workflow. Its 860 ± 30 nm source is the central distinction from the white-light TN420 and TN500, not a cosmetic change of model number.

Choose TN480 when your required procedure calls for the manufacturer-stated ISO 7027 / DIN EN 27027 infrared method. The source is also positioned for appropriate colored samples, but no optical design eliminates every matrix interference.

Apera TN480 product photograph
Portable TN480 configuration. Operate on a level surface; do not hold during measurement.

0-1000 NTU

Instrument range

Method/reporting scope is a separate requirement.

860 ± 30 nm

Infrared LED

Match the required optical method.

Windows records

PC-Link + selectable TruRead

Capacity depends on mode.

Choose your turbidity workflowClick or tap to compare 4 meters + standardsChoose an item below - click or tap here to close

OPTICAL METHOD BEFORE UNIT LABEL

Infrared is a method choice,
not a white-light relabeling.

TN480

Infrared LED, 860 ± 30 nm, with manufacturer-stated ISO 7027 / DIN EN 27027 positioning.

TN420 / TN500

Tungsten white light, 400-600 nm, with manufacturer-stated EPA 180.1 positioning.

Reporting

Optics, geometry, sample matrix and the required method govern interpretation. A unit label is not a universal conversion rule.

Apera positions this infrared design for appropriate colored samples. It is not immune to every interference, and it does not become an EPA white-light meter because a calibration material carries an EPA-related claim. The exact manual describes NTU/FTU display settings; a separately selectable FNU mode is not asserted.

WHY CHOOSE THIS CONFIGURATION

Match the work, not just the feature count.

Keep the optical method explicit

860 ± 30 nm infrared LED and 90° scattering distinguish this setup from a tungsten white-light meter. Match the instrument to the required method before comparing records features.

Review the complete sequence

Selectable TruRead provides individual values, average, minimum and maximum. Changing samples still need consistent mixing, bubble control and timing.

Separate data transfer from charging

Windows PC-Link and USB support reporting. The power specification is four AA alkaline batteries, not a rechargeable lithium pack or an implied USB-charging feature.

A sequence and a record,
with the limits made visible.

TN480 offers Normal measurement and selectable 5/10/15/20-reading TruRead with average, minimum, maximum and individual values. Its manual gives 10 seconds for a Normal value; the TruRead example uses 20-second intervals.

USB connects to the supplied Windows PC-Link workflow for transfer and Excel-document export. Specific current Windows versions, Mac/mobile support and cloud backup are not established; check your IT/driver requirements before deployment.

200 sets is not the capacity
of every recording mode.

Plan storage around the selected sequence, not the largest headline number. The exact manuals give these onboard capacities:

Measurement mode Stored sets
Normal 200
TruRead: 5 readings 100
TruRead: 10 readings 61
TruRead: 15 readings 44
TruRead: 20 readings 34

Mixed use changes the available capacity. Export needed records before clearing the instrument; its clear-all function removes stored records rather than offering a selective archival system.

PC-Link supports a separate live-PC workflow. When that workflow is running, the manuals describe measurements going to the computer rather than being stored in the instrument. This is not cloud backup or a verified electronic-signature/audit-trail system.

READ THE TWO NUMBERS DIFFERENTLY

Display resolution is not accuracy.

0.01 NTU

Low-range display increment

Other ranges use 0.1 and 1 NTU increments.

2% + stray light

Published reading-error specification

Accuracy: ≤±2% of reading plus stray light; stray light ≤0.02 NTU.

Worked specification example: at 1.00 NTU, the 2%-of-reading term alone is 0.02 NTU before the stray-light term. That arithmetic is not a complete method-uncertainty calculation or a promise of ±0.01 NTU measurement accuracy. Vials, sample preparation, calibration and the matrix remain part of the result.

Apera TN480 exact kit case and supplied components
TN480 carrying-case configuration. The readable checklist defines the included items; a computer is not supplied.

Know what arrives
in the carrying case.

TN480 meter; five standards at 0/20/100/400/800 NTU; six numbered sample vials; 10 mL silicone oil; microfiber cloth; USB cable; PC-Link software flash drive; four AA alkaline batteries; battery-cover screwdriver; manual and case.

The complete meter kit includes zero. Replacement nonzero standards are a separate purchase; do not read a replacement-set name as all five calibration solutions.

SAMPLE PREPARATION IS PART OF THE MEASUREMENT

The vial and the timing
can change the answer.

Prepare the sample vial

Use the specified 25 × 60 mm numbered sample vial and approximately 18 mL of sample. Keep the matching cap paired with it and follow the exact rinsing/mixing procedure.

Control bubbles and settling

Fine bubbles can raise readings; settling and light-absorbing material can change them. Keep preparation and waiting times consistent rather than averaging away a method problem.

Protect the optical surfaces

Keep the vial clean and dry, avoid fingerprints/scratches, align its mark and close the cover. Use oil only as instructed and remove excess.

Work on a flat, level surface, not with the meter in your hand. Never pour sample into or wash the optical holder. The IP67 rating is not a direct-immersion measurement method. Turbidity is not a direct TSS mass measurement or a complete water-safety test.

Check zero and the right standards.
Keep nonzero vials sealed.

The kit is supplied with five calibration values and is factory calibrated across its range. The manual permits later calibration with two or more appropriate points bracketing the expected range; local method/QC requirements still govern. Select the correct value because the instrument does not identify a standard automatically.

For low-range work, follow the manual’s zero check, warm-up and numbered-vial guidance. A display increment is not proof that an unchecked zero or contaminated vial is adequate.

The four nonzero AMCO standard vials must stay sealed. Zero-vial opening/refill instructions are different. Do not vigorously shake standards or infer that a generic shelf-life claim permits opening them.

Order infrared replacements,
not look-alike white-light standards.

The exact TN480 replacement set is T200-1 for the four nonzero values. T500-1 is assigned to TN420/TN500 and is not represented as compatible with this instrument. Zero refill is a separate item.

Alternative optical method: TN500 white-light recording meter and Alternative optical method: TN420 white-light meter are different method choices, not equivalent replacements for an infrared requirement.

Watch Apera’s turbidity overview.

The official family tutorial introduces the workflow. Use the exact TN480 manual for its controls, optical method, standards and software. The video is not a model-specific certification or a replacement for sealed-vial instructions.

TN480 buying and workflow questions

Is TN480 the EPA white-light alternative to TN500?

TN480 AI486 uses infrared LED optics and is positioned by Apera for ISO 7027 / DIN EN 27027. TN500 uses tungsten white light. Choose the required method first; sharing a turbidity range does not make the methods interchangeable.

Can NTU and FNU labels be treated as a universal conversion?

No. Instrument optics, geometry and reporting method matter. Source literature uses varying unit labels; the exact manual describes NTU/FTU display settings, not a separately verified FNU selector. Do not convert or substitute methods by changing a label alone.

How many records can TN480 actually hold?

It holds 200 Normal sets, or 100/61/44/34 TruRead sets at 5/10/15/20 readings respectively. Mixed-mode capacity depends on the records saved. Clear-all is irreversible; export needed records first.

Can I use T500-1 standards with TN480?

Do not select them for TN480. The manufacturer assigns T200-1 to TN480/TN400 infrared models and T500-1 to TN500/TN420 white-light models. Matching 20/100/400/800 labels alone is not compatibility evidence.

What power and computer setup does TN480 need?

The meter uses four AA alkaline batteries. Its documented USB/PC-Link reporting workflow is Windows-based with software supplied on a flash drive. No built-in rechargeable-lithium specification, Mac/mobile support or modern Windows-version guarantee is established.

What is supplied with the TN480 meter?

The kit lists five standards at 0/20/100/400/800 NTU, six sample vials, 10 mL silicone oil, microfiber cloth, USB cable, software flash drive, four AA batteries, screwdriver, manual and case. A Windows computer is not included.

Does TruRead replace careful sample preparation?

No. It helps show changing readings and their spread. Use a level surface, clean indexed vials, controlled mixing/timing and bubble handling. Never pour sample into or wash the optical vial holder.

Is the low-range 0.01 NTU display increment the accuracy?

No. Published accuracy is ≤±2% of reading plus stray light, with stray light ≤0.02 NTU. The display increments are 0.01/0.1/1 NTU across ranges. These are not interchangeable specifications or a total method-uncertainty guarantee.

Confirm the method.
Choose the exact configuration.

Review TN480 AI486 and return to the live price and ordering controls. Ask Blue Tiger Scientific to clarify the exact configuration, method or documentation requirements before ordering.

Back to price and ordering ↑
    SpecificationDetail
    BrandApera Instruments
    Model / manufacturer catalogTN480 / AI486
    Optical principleInfrared LED, 860 ± 30 nm; 90° nephelometric scattering
    Method positioningManufacturer-stated ISO 7027 / DIN EN 27027 compliance; not an independent approval guarantee
    Instrument range0-1000 NTU; applicable method range is separate
    Display resolution0.01 / 0.1 / 1 NTU across ranges; not accuracy
    Accuracy≤±2% of reading plus stray light
    Stray light≤0.02 NTU
    Repeatability≤±1% of reading or 0.02 NTU, whichever is greater
    Normal measurementNormal value in 10 seconds per manual; TruRead examples use 20-second reading intervals
    Measurement modesNormal, or selectable 5/10/15/20-reading TruRead with average, minimum, maximum and individual values
    Records200 Normal sets; TruRead capacities depend on selected sequence length
    Power4 AA alkaline batteries; not the lithium battery specified for TN420/TN500
    Calibration values0 / 20 / 100 / 400 / 800 NTU
    Replacement nonzero setT200-1 for the four nonzero infrared standards; do not substitute T500-1 on the basis of shared numbers
    Meter-kit zero standardIncluded; not the same contents as a four-value replacement set
    Sample vial25 mm diameter × 60 mm, high-borosilicate glass with cap
    Sample-test volumeApproximately 18 mL; not a standard-kit fill volume
    DisplayTFT color screen
    Ingress ratingIP67; do not pour into or wash the optical holder
    Operating temperature0-50 °C
    Published meter dimensions90 × 203 × 80 mm
    Published meter mass385 g
    Calibration-solution storage5-30 °C; never freeze; follow expiry label
    Onboard capacity by modeNormal 200; TruRead 5/10/15/20 readings: 100/61/44/34 sets
    PC software / exportWindows PC-Link; USB transfer and Excel-document export; no Mac/mobile/cloud guarantee
    Data features, not certificationDate/time, identifiers and calibration information; no verified electronic-signature or complete historical audit-trail claim

    Ordering Information

    Ordering detailIncluded configuration
    Manufacturer catalogAI486
    Preserved Blue Tiger SKUBTS-AI486
    Order contentsTN480 meter; five standards at 0/20/100/400/800 NTU; six numbered sample vials; 10 mL silicone oil; microfiber cloth; USB cable; PC-Link software flash drive; four AA alkaline batteries; battery-cover screwdriver; manual and case.
    Order unit1 instrument kit

    Warranty Information

    APERA INSTRUMENTS, LLC warrants this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at the option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to the responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS for the instrument from the delivery.

    This limited warranty does NOT cover any issues due to:

    • Accidental damage
    • Improper use
    • Normal wear and tear
    • Transportation
    • Storage
    • Failure to follow the product instructions
    • Unauthorized maintenance, modifications, combination or use with any products, materials, processes, systems or other matter
    • Unauthorized repair
    • External causes such as accidents, abuse, or other actions or events beyond our reasonable control.

    TN480 official documents

    The exact-model manual controls operation and configuration. The family brochure is a supplemental selection reference; it also illustrates other models and accessories. A four-value replacement set is not the complete five-standard meter-kit package. Each card opens a PDF in a new tab.