Apera TN420 Portable White-Light Turbidity Meter

$795.00

    White-light testing. A focused, portable workflow. TN420 AI482 for the defined measurement workflow.

    • White-light, 90° scattering; tungsten filament lamp, 400-600 nm
    • Normal reading, or fixed 10-reading average
    • USB charging; no documented PC-Link export
    • Includes five standards and six sample vials

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

Preparing Quote Cart. Your shopping cart stays unchanged.

APERA INSTRUMENTS / TN420 / AI482

White-light testing.
A focused, portable workflow.

TN420 is the straightforward white-light choice when you need portable turbidity measurements without an electronic records workflow. Its tungsten-lamp optics, numbered vials and fixed 10-reading averaging mode keep the purchase focused on the measurement task.

Choose it when your required method calls for white-light nephelometry and local readings are enough. Choose TN500 instead if dated records, selectable TruRead reports and Windows export are part of the job. Both have the same published basic range and accuracy specification.

Apera TN420 product photograph
Portable TN420 configuration. Operate on a level surface; do not hold during measurement.

0-1000 NTU

Instrument range

Method/reporting scope is a separate requirement.

400-600 nm

Tungsten white light

Match the required optical method.

Fixed average

10-reading averaging

No documented PC export.

Choose your turbidity workflowClick or tap to compare 4 meters + standardsChoose an item below - click or tap here to close

INSTRUMENT RANGE AND METHOD SCOPE ARE DIFFERENT

0-1000 NTU is the meter range.
It is not the undiluted EPA method range.

Apera describes TN420 as EPA 180.1 compliant. EPA 180.1 Rev. 2.0 covers 0-40 NTU and directs dilution for samples above 40 NTU. A broader instrument scale helps characterize a sample; it does not remove the method’s preparation, checks or reporting requirements.

Instrument positioning

Manufacturer-stated method compliance is not a model-specific EPA approval certificate.

Procedure and quality control

Follow the applicable method and program, including required verification, dilution and recordkeeping. A general weekly-check suggestion does not replace required method-specific checks.

WHY CHOOSE THIS CONFIGURATION

Match the work, not just the feature count.

Choose the method first

Tungsten white-light nephelometry is a different optical choice from the infrared TN480. A unit label alone does not make the methods interchangeable.

Average a defined sequence

The average mode takes 10 readings at 20-second intervals. It does not offer the selectable 5/10/15/20-reading report workflow of TN500.

Keep the setup portable

Recharge by USB and use the supplied case for transport. During a reading, place the instrument on a flat, level surface rather than holding it in your hand.

A defined average,
not a stored TruRead report.

Normal mode takes a reading in 20 seconds. Average mode runs 10 consecutive readings at 20-second intervals and reports the average, about 200 seconds for the sequence before allowing for preparation and other checks.

There is no documented TN420 onboard logging/recall or PC-Link export function. USB charging from an adapter or computer should not be mistaken for electronic result transfer. If reportable sequences or dated records are needed, compare Alternative recording workflow: TN500.

READ THE TWO NUMBERS DIFFERENTLY

Display resolution is not accuracy.

0.01 NTU

Low-range display increment

Other ranges use 0.1 and 1 NTU increments.

2% + stray light

Published reading-error specification

Accuracy: ≤±2% of reading plus stray light; stray light ≤0.02 NTU.

Worked specification example: at 1.00 NTU, the 2%-of-reading term alone is 0.02 NTU before the stray-light term. That arithmetic is not a complete method-uncertainty calculation or a promise of ±0.01 NTU measurement accuracy. Vials, sample preparation, calibration and the matrix remain part of the result.

Apera TN420 exact kit case and supplied components
TN420 carrying-case configuration. Its charger and USB cable support charging, not a documented PC-Link function.

Know what arrives
in the carrying case.

TN420 meter; five standards at 0/20/100/400/800 NTU; six numbered sample vials; 10 mL silicone oil; microfiber cloth; 5 V/1 A adapter; USB cable; manual and case.

The complete meter kit includes zero. Replacement nonzero standards are a separate purchase; do not read a replacement-set name as all five calibration solutions.

SAMPLE PREPARATION IS PART OF THE MEASUREMENT

The vial and the timing
can change the answer.

Prepare the sample vial

Use the specified 25 × 60 mm numbered sample vial and approximately 18 mL of sample. Keep the matching cap paired with it and follow the exact rinsing/mixing procedure.

Control bubbles and settling

Fine bubbles can raise readings; settling and light-absorbing material can change them. Keep preparation and waiting times consistent rather than averaging away a method problem.

Protect the optical surfaces

Keep the vial clean and dry, avoid fingerprints/scratches, align its mark and close the cover. Use oil only as instructed and remove excess.

Work on a flat, level surface, not with the meter in your hand. Never pour sample into or wash the optical holder. The IP67 rating is not a direct-immersion measurement method. Turbidity is not a direct TSS mass measurement or a complete water-safety test.

Check zero and the right standards.
Keep nonzero vials sealed.

The kit is supplied with five calibration values and is factory calibrated across its range. The manual permits later calibration with two or more appropriate points bracketing the expected range; local method/QC requirements still govern. Select the correct value because the instrument does not identify a standard automatically.

For low-range work, follow the manual’s zero check, warm-up and numbered-vial guidance. A display increment is not proof that an unchecked zero or contaminated vial is adequate.

The four nonzero AMCO standard vials must stay sealed. Zero-vial opening/refill instructions are different. Do not vigorously shake standards or infer that a generic shelf-life claim permits opening them.

Match replacement standards
to this white-light model.

Compatible accessory: T500-1 sealed AMCO nonzero standards is explicitly listed for TN420 and the other white-light model. Its four values are 20/100/400/800 NTU; it does not supply a meter or zero standard.

The infrared TN480 specifies T200-1 instead. Shared concentration labels do not establish universal cross-model or cross-brand compatibility.

Watch Apera’s turbidity overview.

The official family tutorial introduces the workflow. Use the exact TN420 manual for its controls, optical method, standards and software. The video is not a model-specific certification or a replacement for sealed-vial instructions.

TN420 buying and workflow questions

How is TN420 different from TN500?

TN420 AI482 and TN500 AI483 share a 0-1000 NTU range and the same published basic accuracy. TN420 has fixed 10-reading averaging and no documented record export; TN500 adds selectable TruRead, dated records and Windows PC-Link.

Is TN420 an EPA-approved meter?

Apera describes TN420 as EPA 180.1 compliant. That is not a model-specific EPA approval or a guarantee that a laboratory procedure meets every program requirement. Use the applicable method, standards, quality controls and reporting rules.

Does the 0-1000 NTU instrument range equal the EPA method range?

No. EPA 180.1 Rev. 2.0 covers 0-40 NTU and directs dilution for samples above 40 NTU. The meter’s broader display range is not a claim of undiluted regulatory-method coverage across 0-1000 NTU.

Does 0.01 NTU mean accuracy of ±0.01 NTU?

No. It is the low-range display increment. Published accuracy is ≤±2% of reading plus stray light; stray light is specified at ≤0.02 NTU. Sample preparation and method uncertainty are separate.

Can the TN420 USB port export results?

The exact TN420 manual documents USB charging, not PC-Link export or an onboard record-recall function. Do not buy TN420 expecting the electronic reporting supplied with TN500 or TN480.

Are standards and sample vials included?

Yes. The meter kit lists standards at 0, 20, 100, 400 and 800 NTU, six sample vials, silicone oil, cloth, charger, USB cable, manual and case. Replacement T500-1 contains only the four nonzero standards, not zero.

Can I hold the meter or pour sample directly into it?

No. Portable does not mean handheld during a reading: use a level surface. Sample goes in the capped sample vial, not directly into the optical holder. Never wash the holder or treat IP67 as a direct-immersion measurement method.

Does a turbidity reading establish TSS or drinking-water safety?

No. Turbidity is an optical scattering measurement, not a direct total-suspended-solids mass measurement or a complete potability test. A single NTU result does not establish water safety.

Confirm the method.
Choose the exact configuration.

Review TN420 AI482 and return to the live price and ordering controls. Ask Blue Tiger Scientific to clarify the exact configuration, method or documentation requirements before ordering.

Back to price and ordering ↑
    SpecificationDetail
    BrandApera Instruments
    Model / manufacturer catalogTN420 / AI482
    Optical principleWhite-light, 90° scattering; tungsten filament lamp, 400-600 nm
    Method positioningManufacturer-stated EPA 180.1 compliance; not an independent approval guarantee
    Instrument range0-1000 NTU; applicable method range is separate
    Display resolution0.01 / 0.1 / 1 NTU across ranges; not accuracy
    Accuracy≤±2% of reading plus stray light
    Stray light≤0.02 NTU
    Repeatability≤±1% of reading or 0.02 NTU, whichever is greater
    Normal measurement20 seconds per Normal reading; preparation and quality-control time are additional
    Measurement modesNormal reading, or fixed 10-reading average
    RecordsNo onboard storage, recall or PC export feature documented in the current exact-model manual
    Power3.7 V rechargeable lithium battery; USB charging
    Calibration values0 / 20 / 100 / 400 / 800 NTU
    Replacement nonzero setT500-1 / AI4205 for the four nonzero white-light standards; zero refill is separate
    Meter-kit zero standardIncluded; not the same contents as a four-value replacement set
    Sample vial25 mm diameter × 60 mm, high-borosilicate glass with cap
    Sample-test volumeApproximately 18 mL; not a standard-kit fill volume
    DisplayTFT color screen
    Ingress ratingIP67; do not pour into or wash the optical holder
    Operating temperature0-50 °C
    Published meter dimensions90 × 203 × 80 mm
    Published meter mass385 g
    Calibration-solution storage5-30 °C; never freeze; follow expiry label
    USB scopeCharging documented; no PC-Link export feature established

    Ordering Information

    Ordering detailIncluded configuration
    Manufacturer catalogAI482
    Preserved Blue Tiger SKUBTS-AI482
    Order contentsTN420 meter; five standards at 0/20/100/400/800 NTU; six numbered sample vials; 10 mL silicone oil; microfiber cloth; 5 V/1 A adapter; USB cable; manual and case.
    Order unit1 instrument kit

    Warranty Information

    APERA INSTRUMENTS, LLC warrants this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at the option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to the responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS for the instrument from the delivery.

    This limited warranty does NOT cover any issues due to:

    • Accidental damage
    • Improper use
    • Normal wear and tear
    • Transportation
    • Storage
    • Failure to follow the product instructions
    • Unauthorized maintenance, modifications, combination or use with any products, materials, processes, systems or other matter
    • Unauthorized repair
    • External causes such as accidents, abuse, or other actions or events beyond our reasonable control.

    TN420 official documents

    The exact-model manual controls operation and configuration. The family brochure is a supplemental selection reference; it also illustrates other models and accessories. A four-value replacement set is not the complete five-standard meter-kit package. Each card opens a PDF in a new tab.