When a glass thermometer is used over a sample, the coating and the immersion mark answer different questions. The FEP sleeve is designed to retain broken glass and fill if the instrument breaks. The immersion type tells you how much of the stem belongs in the medium for a reading.

For a −20 to 110°C bath, 80707E and 80907E both have 1°C marks. Use the total-immersion 80707E when the procedure immerses the bulb and liquid column up to the reading. Choose 80907E when the bath surface must meet the partial-immersion line.
The 30 to 220°F models are 15.38 inches long, compared with 12 inches for the other ranges. Check clearance around the vessel and clamp before choosing 80704E or 80904E. Overall length is not a specified immersion depth.
The coating is designed to retain the glass fragments and liquid fill. It does not make the instrument unbreakable. If damage is suspected, remove it from service and follow your facility’s procedure for handling the instrument and any affected sample.
Not on the basis of range alone. Total immersion and partial immersion place different lengths of the liquid column in the medium. Match the designation in the method; for a partial-immersion instrument, look for the line on the actual stem.
80706E (total) and 80906E (partial) reach −35°C, but stop at 50°C. The −20 to 110°C and −20 to 150°C pairs reach higher temperatures, but neither covers the colder interval.
Choose 80704E for total immersion or 80904E for partial immersion. Both have 2°F divisions and a 15.38-inch overall length. If clearance is limited, do not assume that a shorter model with an overlapping range meets the procedure.
No. The catalog number specifies one scale: 1°C divisions on the Celsius models or 2°F divisions on the Fahrenheit models. Select the units your procedure records. The family photograph is not a scale reference for every model.
The serial number identifies the individual instrument in your equipment log. It does not, by itself, supply calibration results or a certificate. Confirm any required certificate, test points or traceability before ordering.
Each selected catalog number supplies one FEP-encapsulated glass thermometer. These E-suffix models have no metal armor. If your procedure specifies a guard, do not substitute this family simply because the range and graduations match.