Hanna HI96779 Rapid Chlorine Dioxide Reagents, 100/300 Tests

$74.79

    Keep rapid chlorine dioxide checks on the matched DPD method. Use the exact instrument method and reporting basis; bench equipment supplied separately.

    • Rapid DPD method, 0.00–2.00 mg/L as ClO₂
    • Mandatory HI96779A-0 glycine masking followed by HI96779B-0
    • Nominal 100/300-test refills; collect cleanly and analyze immediately

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

Keep rapid chlorine dioxide checks on the matched DPD method.

Bring a shorter, clearly defined color-development sequence to chlorine dioxide monitoring in drinking, tap and treated water. HI96779 combines its specified glycine masking reagent A with DPD reagent B for the rapid photometer method. Choose a nominal 100- or 300-test refill while keeping sample handling, reagent order and reporting as ClO₂ consistent.

Official Hanna HI96779 reagent packaging reference
Official Hanna HI96779 reagent packaging reference. Photo is illustrative, not a bottle/packet inventory for either pack; printed lot, expiry and any container markings do not specify supplied stock.

Match the refill to the complete method

Match the Chlorine Dioxide, Rapid Method program—not a conventional chlorine dioxide or chlorine program. The documented range is 0.00–2.00 mg/L as ClO₂ on HI83300, HI83305, HI83326 and HI83399. Mandatory HI96779A-0 masking reduces free-chlorine response, but does not remove every oxidant interference.

Two components with distinct jobs

Chlorine dioxide reacts with DPD to produce the measured pink tint. HI96779A-0 is the glycine masking reagent: it inhibits free-chlorine response without replacing the DPD indicator in HI96779B-0. The specified method requires reagent A in the sequence; it is not an optional helper added only when chlorine has been detected.

Collect cleanly and analyze immediately

Use a clean glass collection bottle and analyze without delay. Chlorine dioxide is a strong oxidizer and is unstable in water; delayed testing can change the sample before the photometer measures it. The documented applications are drinking water, tap water and treated water. The method does not establish universal suitability for every industrial, colored or oxidant-rich matrix.

Keep the masking step before zeroing

Fill a clean cuvette with 10 mL sample, add 5 drops of HI96779A-0, replace stopper and cap, and shake gently for 30 seconds. Wait another 30 seconds, then zero. Add one packet of HI96779B-0 to that cuvette, cap and shake gently for 20 seconds, then wait 1 minute and read. Follow the exact model manual; the one-minute development time is not the total sampling-to-result time.

Account for oxidants and matrix effects

The manual flags acidity, alkalinity, flocculating agents, hardness, inorganic and organic chloramines, manganese, metals, monochloramine, oxidized chromium/manganese, ozone and peroxides. Chlorine above 5 mg/L, bromine above 0.1 mg/L, high buffering and extreme sample pH are also interference conditions. Glycine masking does not make the method selective against all oxidants; assess the actual matrix and reporting need.

Do not transfer another chlorine dioxide workflow

HI93738 uses the conventional four-liquid chlorophenol-red method and is not the same reagent system. HI93703-52 ozone-method glycine powder is not the specified HI96779A-0 masking reagent; exact official code equivalence has not been established. Neither a shared chemical name nor nearby accessory placement supports substitution or an extra helper addition.

Plan the refill and keep its safety record

The nominal 100- and 300-test labels identify analytical refill capacity, not independently verified bottle counts, fill volumes or packet inventory. Provide the compatible meter, clean cuvette/cap/stopper and clean glass collection bottle separately. Follow both actual US component SDS and current labels for PPE, storage and waste handling, and retain lot/expiry identification. No fixed shelf life, dispatch promise or air-transport permission is inferred.

Frequently Asked Questions

What changes between the 100- and 300-test options?

HI96779-01 is labeled for 100 tests and HI96779-03 for 300 tests. Both use the same rapid DPD method with HI96779A-0 followed by HI96779B-0. These are nominal test counts, not a claim about physical bottle/packet counts, fill volumes or guaranteed reportable samples after QC and repeats.

Is reagent A optional if I expect no free chlorine?

No. The documented procedure always adds 5 drops of HI96779A-0 before shaking, waiting and zeroing. Its glycine masks the response of free chlorine. Do not skip that step or move it after zeroing based on an assumption about the sample.

Which instruments have the exact rapid method?

HI83300, HI83305, HI83326 and HI83399 manuals document the HI96779 A/B rapid chlorine dioxide procedure. Select that exact program and follow the model-specific operating instructions. Compatibility is not inferred from generic cuvette fit or an accessory located near a reagent listing.

How quickly can the sample be read?

Collect in clean glass and analyze immediately. For 10 mL sample: add 5 drops A, shake gently 30 seconds, wait 30 seconds and zero; then add one packet B, shake gently 20 seconds, wait 1 minute and read. The one-minute final wait excludes collection, masking, zeroing and handling time.

What does the result represent?

The range is 0.00–2.00 mg/L as chlorine dioxide, ClO₂, with 0.01 mg/L resolution. It is not a result as chlorine Cl₂, free chlorine, ozone or total oxidants. The method is a DPD adaptation of Standard Methods 18th edition, 4500 ClO₂ D; record the analyte and reporting basis explicitly.

Can I substitute HI93738 or HI93703-52?

No substitution is established. HI93738 is the conventional four-liquid chlorophenol-red chemistry, not this rapid DPD system. HI93703-52 is an ozone-method glycine powder, not the method-specified HI96779A-0. Exact official code equivalence has not been established, so neither replacement nor an extra cross-sold helper is recommended.

Does glycine eliminate all oxidant interference?

No. The method still flags other oxidants and matrix conditions, including ozone, peroxides, chloramines, oxidized metals, acidity/alkalinity, high buffering and extreme pH. Chlorine above 5 mg/L and bromine above 0.1 mg/L are specified interference conditions. Assess sample suitability under the full method, not just the masking step.

Where are handling, storage and expiry instructions?

Use the actual HI96779A-0 and HI96779B-0 US SDS in the document tab together with the current supplied labels. The instrument manuals are procedure references, not standalone consumable manuals. Follow their PPE, handling, storage and waste directions, retain lot/expiry identification and do not use the illustrative photo to infer supplied expiry or a fixed remaining shelf life.

    Product Specifications

    SpecificationDetail
    Parameter / report basisChlorine dioxide, mg/L as ClO₂; not Cl₂ or total oxidants
    MethodRapid DPD adaptation, Standard Methods 18th ed., 4500 ClO₂ D
    Range / resolution0.00–2.00 mg/L as ClO₂ / 0.01 mg/L
    Instrument-method accuracy±0.10 mg/L ±5% of reading at 25 °C; method specification, not a reagent-only guarantee
    Optical method525 nm instrument method
    Procedure-proven instrumentsHI83300, HI83305, HI83326, HI83399 — Chlorine Dioxide, Rapid Method
    Mandatory masking componentHI96779A-0: 5 drops per 10 mL sample, before zeroing
    Indicator componentHI96779B-0: 1 packet after masking and zeroing
    Timing before zeroAfter A: shake gently 30 s, then wait 30 s
    Timing before readAfter B: shake gently 20 s, then wait 1 min
    Sampling / matricesClean glass bottle; analyze immediately. Drinking, tap and treated water
    Threshold interferencesChlorine >5 mg/L; bromine >0.1 mg/L
    Other interferencesAcidity, alkalinity, flocculants, hardness, chloramines, manganese/metals, oxidized Cr/Mn, ozone, peroxides; high buffering/extreme pH
    Chemistry boundaryNot conventional HI93738; no established equivalence between HI96779A-0 and HI93703-52
    Pack-count basisNominal 100 or 300 tests; physical bottle/packet inventory and fill volumes not independently enumerated
    Required separatelyCompatible photometer, clean 10 mL cuvette/cap/stopper, clean glass sampling bottle, accurate timing and SDS-prescribed PPE

    Ordering Information

    Catalog numberReagent pack
    HI96779-01100 nominal tests; one sellable reagent pack
    HI96779-03300 nominal tests; one sellable reagent pack

    Shipping, Return, and Warranty Policy

    At Blue Tiger Scientific, we are proud to distribute authentic Hanna Instruments products. The following policies apply to all purchases made through our store.

    Shipping Policy

    • All orders are shipped via FedEx from Hanna Instruments' fulfillment center in Smithfield, Rhode Island.
    • FedEx Ground is the standard method; expedited options (2-Day, Overnight) are available for an additional cost.
    • Hanna Instruments only ships to residential or business addresses within the United States (no PO Boxes).
    • Orders placed after business hours, weekends, or holidays will ship the next business day, pending product and carrier availability.
    • Shipping timelines may vary depending on product availability, order size, and carrier logistics.
    • Blue Tiger Scientific and Hanna Instruments are not responsible for carrier delays, weather-related disruptions, or other unforeseen delivery issues once the order has shipped.

    Return Policy

    • All return requests must be coordinated directly with Blue Tiger Scientific before a product is sent back.
    • Products must have a Return Merchandise Authorization (RMA) number issued by Blue Tiger Scientific before being returned.
    • Products must be returned in new or like-new condition, with the original packaging and proof of purchase (such as an invoice or packing slip).
      Customers are responsible for return shipping costs unless otherwise stated.
    • Unauthorized returns (without an RMA) will be refused and returned at the customer's expense.
      Returns are subject to inspection and approval. Products not in acceptable condition may be denied credit.
    • Some products, such as custom-built, special order, and hazardous materials, may not be eligible for return.

    Warranty Policy

    • Hanna Instruments' products are covered by their Limited Warranty against defects in materials and workmanship under normal use.
    • All warranty claims and services are handled directly by Hanna Instruments.
    • Warranty periods vary by product. If unspecified, the standard warranty term is 18 months from the Hanna Instruments factory ship date.
    • If you experience an issue believed to be covered under warranty, please contact Hanna Instruments for guidance on how to submit your warranty request to Hanna Instruments.
    • Customers must follow Hanna Instruments' warranty submission process, including providing a claim number, product model and serial number, and a detailed description of the issue.
    • Hanna Instruments will determine whether products are repaired, replaced, or otherwise serviced under warranty.
    • Note: Warranty coverage does not apply to damage caused by misuse, accidents, unauthorized repairs, or improper storage and handling.

    Important Notes

    • Some Hanna Instruments products are custom-made or built to order. Shipping will occur once the product is ready.
    • In the event of backorders, you may request partial shipments.
    • Delays beyond our control (e.g., material shortages, import delays, or carrier issues) may occasionally occur.

Questions about a product? Ask us here.

email us here: info@bluetigerscientific.com or live chat

Your Carts

Which would you like to open?

Shopping Cart

0 items

Review items for online purchase.

View Cart

Quote Cart

0 items

Review items and request a quote.

View Quote Cart

Your two carts stay separate. Nothing is moved or combined.