Hanna HI95771 Total Chlorine Ultra High Range Reagents, 100/300 Tests Photometer Reagents Hanna Instruments
Hanna HI95771 Total Chlorine Ultra High Range Reagents, 100/300 Tests Photometer Reagents Hanna Instruments

Hanna HI95771 Total Chlorine Ultra High Range Reagents, 100/300 Tests

$40.69

    Keep total-chlorine UHR measurements supplied with matched A/B chemistry Keep HI97771 water-sample testing supplied with the complete matched total chlorine ultra high range reagent set.

    • HI95771-01: 100 tests; HI95771-03: 300 tests
    • HI97771: 0–500 mg/L as chlorine (Cl2), total chlorine UHR
    • Reagents only; meter, cuvettes and CAL Check standards separate

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

HANNA PHOTOMETER REAGENTS · HI95771

Keep total-chlorine UHR measurements supplied with matched A/B chemistry

Choose Hanna HI95771 to replenish the two-powder reagent system for total chlorine at ultra high range. The HI97771 Total Chlorine UHR method measures 0–500 mg/L as Cl2. Nominal 100-test and 300-test packs support the same method, so replenishment can follow workload without changing analytical chemistry.

Official Hanna HI95771 reagent set reference photo
Manufacturer reference photo identifies the two powder components. Pictured lot/expiry markings are illustrative, not the supplied lot or an inventory count.

Match the method, not just the element name

HI97771: 0–500 mg/L as chlorine (Cl2), total chlorine UHR. Two powder components: HI95771A-0 and HI95771B-0. Not ultra low range and not free-chlorine UHR chemistry.

What the selected pack supplies

HI95771-01 and HI95771-03 supply complete matched total-chlorine UHR reagent sets with manufacturer-nominal capacities of 100 and 300 tests. A determination uses both powder components; nominal capacity is not an independently verified count of the physical packets in the retail box.

Where this method fits

Reserve this family for the Total Chlorine UHR method. The HI97771 also offers free-chlorine measurements at 0.00–5.00 mg/L using different chemistry; that capability does not make HI95771 a free-chlorine reagent. HI83300 and HI83399 also document the HI95771 reagent family; follow their instrument-specific procedures.

The HI97771 workflow at a glance

Select Total Chlorine UHR on HI97771 and zero with 10 mL of unreacted sample. Remove the cuvette, add one packet each of HI95771A-0 and HI95771B-0, replace stopper and cap, and shake gently for 20 seconds. Reinsert as directed and use the one-minute read countdown. Follow the linked current manual for complete operation; do not borrow the free-chlorine procedure.

Check the sample matrix

Bromine, chlorine dioxide, chromium, iodine, oxidized manganese and ozone may interfere. Review the sample’s oxidant matrix before interpreting a result as total chlorine. The stated 0–500 mg/L range is the HI97771 total-chlorine UHR method range, not a promise of suitability for every sample matrix.

Choose a pack around your workload

HI95771-01 supplies 100 tests; HI95771-03 supplies 300 tests. The larger pack supplies three times the determinations, not a different measuring range. One order quantity means one complete selected pack. Choose by testing frequency and the lot/expiry markings on supplied stock.

What to have beside the reagent set

Supply the compatible photometer, cuvettes with stoppers and caps, and the measuring/dispensing accessories required by the method separately. CAL Check standards are separate instrument-verification materials, not replacements for sample reagents.

Handling before you begin

Review both current US component SDS documents for PPE, handling, storage, spill response and disposal. Keep the two powders identified and use both components for the method. Check the lot and expiry on supplied stock rather than the older illustrative photo. Hanna marks these offers as Dangerous Goods, with ground shipment only.

Use the linked HI97771 instrument manual as the method reference for the complete procedure, precautions and instrument operation. This is not a separate reagent instruction manual.

Frequently Asked Questions

Should I choose 100 or 300 tests?

HI95771-01 supplies 100 tests; HI95771-03 supplies 300 tests. Both are complete reagent sets for the same method. Choose by expected workload and marked expiry dates, not by a different analytical range.

Is UHR the same as ultra low range?

No. UHR means ultra high range here: 0–500 mg/L as Cl2 for total chlorine on HI97771. It is not an ultra-low-range reagent family.

Does this also measure free chlorine on HI97771?

No. HI95771 is for Total Chlorine UHR only. HI97771’s free-chlorine range is 0.00–5.00 mg/L and uses different reagents. Do not infer free-chlorine UHR capability from the instrument’s combined product name.

Are both reagents powders?

Yes. The documented method uses one packet of HI95771A-0 and one packet of HI95771B-0 per determination. Both components are required; this is not a liquid A/B reagent system.

What mixing and timing does the primary method use?

After zeroing 10 mL of sample, add both specified powders, replace the stopper and cap, and shake gently for 20 seconds. Use the one-minute read countdown and follow the current HI97771 manual for the complete sequence.

Which photometers are documented for HI95771?

Both packs are documented for HI97771, HI83300, HI83399. The procedure and range on this page use HI97771 as the primary reference; follow the exact instrument method when using another model.

Which sample conditions should I check first?

Bromine, chlorine dioxide, chromium, iodine, oxidized manganese and ozone may interfere. Review the sample’s oxidant matrix before interpreting a result as total chlorine. The stated 0–500 mg/L range is the HI97771 total-chlorine UHR method range, not a promise of suitability for every sample matrix.

Are the instrument, cuvettes and CAL Check standards included?

No. These are reagent-only packs. Provide the photometer, cuvettes with stoppers and caps, and the method-required sampling and dispensing accessories separately. CAL Check standards are separate instrument-validation materials, not test reagents.

    Product Specifications

    SpecificationDetail
    ParameterTotal Chlorine Ultra High Range
    Reagent formatTwo-component powder reagent system
    MethodAdaptation of Standard Methods, 20th Edition, 4500-Cl
    Primary instrumentHI97771
    Range and reporting basis0–500 mg/L as chlorine (Cl2), total chlorine UHR
    Documented instrumentsHI97771, HI83300, HI83399
    Reagent doseOne packet HI95771A-0 plus one packet HI95771B-0 per determination
    Sample and zero preparation10 mL unreacted sample used for zero; then add both powders
    Timing on HI97771Shake gently for 20 seconds; 1-minute read countdown
    MatrixBromine, chlorine dioxide, chromium, iodine, oxidized manganese and ozone may interfere. Review the sample’s oxidant matrix before interpreting a result as total chlorine. The stated 0–500 mg/L range is the HI97771 total-chlorine UHR method range, not a promise of suitability for every sample matrix.
    Pack contentsHI95771-01 and HI95771-03 supply complete matched total-chlorine UHR reagent sets with manufacturer-nominal capacities of 100 and 300 tests. A determination uses both powder components; nominal capacity is not an independently verified count of the physical packets in the retail box.
    Separate requirementsCompatible meter, cuvettes, stoppers, caps, method-required sample/dispensing accessories and CAL Check standards
    Shipping limitationDangerous Goods; ground shipment only

    Ordering Information

    Catalog numberComplete reagent pack
    HI95771-01100 tests — complete total chlorine ultra high range reagent set
    HI95771-03300 tests — complete total chlorine ultra high range reagent set

    Shipping, Return, and Warranty Policy

    At Blue Tiger Scientific, we are proud to distribute authentic Hanna Instruments products. The following policies apply to all purchases made through our store.

    Shipping Policy

    • All orders are shipped via FedEx from Hanna Instruments' fulfillment center in Smithfield, Rhode Island.
    • FedEx Ground is the standard method; expedited options (2-Day, Overnight) are available for an additional cost.
    • Hanna Instruments only ships to residential or business addresses within the United States (no PO Boxes).
    • Orders placed after business hours, weekends, or holidays will ship the next business day, pending product and carrier availability.
    • Shipping timelines may vary depending on product availability, order size, and carrier logistics.
    • Blue Tiger Scientific and Hanna Instruments are not responsible for carrier delays, weather-related disruptions, or other unforeseen delivery issues once the order has shipped.

    Return Policy

    • All return requests must be coordinated directly with Blue Tiger Scientific before a product is sent back.
    • Products must have a Return Merchandise Authorization (RMA) number issued by Blue Tiger Scientific before being returned.
    • Products must be returned in new or like-new condition, with the original packaging and proof of purchase (such as an invoice or packing slip).
      Customers are responsible for return shipping costs unless otherwise stated.
    • Unauthorized returns (without an RMA) will be refused and returned at the customer's expense.
      Returns are subject to inspection and approval. Products not in acceptable condition may be denied credit.
    • Some products, such as custom-built, special order, and hazardous materials, may not be eligible for return.

    Warranty Policy

    • Hanna Instruments' products are covered by their Limited Warranty against defects in materials and workmanship under normal use.
    • All warranty claims and services are handled directly by Hanna Instruments.
    • Warranty periods vary by product. If unspecified, the standard warranty term is 18 months from the Hanna Instruments factory ship date.
    • If you experience an issue believed to be covered under warranty, please contact Hanna Instruments for guidance on how to submit your warranty request to Hanna Instruments.
    • Customers must follow Hanna Instruments' warranty submission process, including providing a claim number, product model and serial number, and a detailed description of the issue.
    • Hanna Instruments will determine whether products are repaired, replaced, or otherwise serviced under warranty.
    • Note: Warranty coverage does not apply to damage caused by misuse, accidents, unauthorized repairs, or improper storage and handling.

    Important Notes

    • Some Hanna Instruments products are custom-made or built to order. Shipping will occur once the product is ready.
    • In the event of backorders, you may request partial shipments.
    • Delays beyond our control (e.g., material shortages, import delays, or carrier issues) may occasionally occur.

Questions about a product? Ask us here.

email us here: info@bluetigerscientific.com or live chat

Your Carts

Which would you like to open?

Shopping Cart

0 items

Review items for online purchase.

View Cart

Quote Cart

0 items

Review items and request a quote.

View Quote Cart

Your two carts stay separate. Nothing is moved or combined.