Hanna HI93750 Potassium Reagents, 100/300 Tests Photometer Reagents Hanna Instruments
Hanna HI93750 Potassium Reagents, 100/300 Tests Photometer Reagents Hanna Instruments

Hanna HI93750 Potassium Reagents, 100/300 Tests

$186.99

    Two potassium ranges. One matched reagent set. Keep HI97750 water-sample testing supplied with the complete matched potassium reagent set.

    • HI93750-01: 100 tests; HI93750-03: 300 tests
    • HI97750: LR 0.0 to 20.0 mg/L as K; MR 10 to 100 mg/L as K
    • Reagents only; meter, cuvettes and CAL Check standards separate

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

HANNA PHOTOMETER REAGENTS · HI93750

Two potassium ranges. One matched reagent set.

Replenish the liquid-and-powder chemistry for both HI97750 potassium ranges. The same 100- or 300-test set serves low-range testing and the documented diluted medium-range workflow—there is no separate MR reagent pack to select.

Official Hanna HI93750 generic bottle reference image
Official Hanna potassium-page bottle reference image, not a photograph of the complete A/B contents. Printed lot and expiry markings are illustrative, not supplied-stock dates.

Match the method, not just the element name

HI97750: LR 0.0 to 20.0 mg/L as K; MR 10 to 100 mg/L as K. The turbidimetric tetraphenylborate method intentionally develops turbidity. On HI97750, the LR range is 0.0 to 20.0 mg/L as K and MR is 10 to 100 mg/L as K. The instrument can also display potassium oxide (K₂O); the ranges quoted here are as K, not as K₂O.

What the selected pack supplies

A complete potassium set containing liquid reagent A and powder reagent B for 100 or 300 determinations. Both HI97750 ranges use 6 drops A and one B packet. HI93703-51 dispersing agent is not part of the documented HI97750 procedure; no additional dispersing step is required.

Where this method fits

A practical refill for water and nutrient-monitoring programs that specify potassium concentration. Use the instrument’s selected range and matching dilution procedure. The supplier’s broader reagent-use range is not the HI97750 range; do not extend this instrument to 200 mg/L as K.

The HI97750 workflow at a glance

Select Potassium LR and fill the cuvette with 10 mL of sample. Add 6 drops of A, stopper and cap, swirl and zero with the cap notch aligned. Add one packet of B, stopper, cap and shake gently for 1 minute. Use the 3-minute countdown; after it has passed, invert 5 times, return the aligned cuvette and press Read twice. For MR, select Potassium MR and first dilute exactly 20 mL of sample to 100 mL in a volumetric flask with deionized water; then run the same additions on 10 mL of that dilution.

Check the sample matrix

The manual lists chloride above 12000 mg/L, calcium above 10000 mg/L as CaCO₃, magnesium above 8000 mg/L as CaCO₃, sodium above 8000 mg/L and ammonium above 10 mg/L as potential interferences. Check the sample matrix rather than assuming unrestricted brine or seawater suitability.

Choose a pack around your workload

HI93750-01 supplies 100 tests; HI93750-03 supplies 300 tests. The larger pack supplies three times the determinations, not a different measuring range. One order quantity means one complete selected pack. Choose by testing frequency and the lot/expiry markings on supplied stock.

What to have beside the reagent set

Supply the compatible photometer, cuvettes with stoppers and caps, and the measuring/dispensing accessories required by the method separately. A volumetric flask and deionized water are required for potassium MR dilution. CAL Check standards are separate instrument-verification materials, not replacements for sample reagents.

Handling before you begin

Review both US SDS before use. Reagent A is flammable and carries serious toxicity, genetic and cancer hazards; B carries ingestion, inhalation and serious eye-damage hazards. Apply the specified controls and disposal practices. Hanna flags both pack sizes as Dangerous Goods for ground shipment only.

Use the linked HI97750 instrument manual as the method reference for the complete procedure, precautions and instrument operation. This is not a separate reagent instruction manual.

Frequently Asked Questions

Should I choose 100 or 300 tests?

HI93750-01 supplies 100 tests; HI93750-03 supplies 300 tests. Both are complete reagent sets for the same method. Choose by expected workload and marked expiry dates, not by a different analytical range.

Do LR and MR need different reagent packs?

No. Both HI97750 methods list the same HI93750-01 and HI93750-03 sets, using 6 drops of A and one packet of B. MR differs by its preliminary sample dilution and the range selected on the meter.

What dilution does medium range require?

Accurately add 20 mL of sample to a 100 mL volumetric flask and bring to volume with deionized water. Select Potassium MR, then use 10 mL of that prepared dilution for the documented A/B procedure.

Do I add HI93703-51 dispersing agent?

No dispersing-agent step appears in the HI97750 LR or MR method. The required reagents are only HI93750A-0 and HI93750B-0. Do not import a dispersing step from a different chemistry.

When should I remix before reading?

After adding B, shake gently for 1 minute, then wait the 3-minute countdown. Invert the cuvette 5 times after that wait, reinsert it with the cap notch aligned, and press Read twice for the result. The quoted HI97750 ranges are as K, not the optional K₂O display.

Which photometers are documented for HI93750?

Both packs are documented for HI97750. The procedure and range on this page use HI97750 as the primary reference; follow the exact instrument method when using another model.

Which sample conditions should I check first?

The manual lists chloride above 12000 mg/L, calcium above 10000 mg/L as CaCO₃, magnesium above 8000 mg/L as CaCO₃, sodium above 8000 mg/L and ammonium above 10 mg/L as potential interferences. Check the sample matrix rather than assuming unrestricted brine or seawater suitability.

Are the instrument, cuvettes and CAL Check standards included?

No. These are reagent-only packs. Provide the photometer, cuvettes with stoppers and caps, and the method-required sampling and dispensing accessories separately. CAL Check standards are separate instrument-validation materials, not test reagents.

    Product Specifications

    SpecificationDetail
    ParameterPotassium
    Reagent formatLiquid HI93750A-0 reagent A and powder HI93750B-0 reagent B
    MethodAdaptation of the Turbidimetric Tetraphenylborate Method
    Primary instrumentHI97750
    Range and reporting basisLR 0.0 to 20.0 mg/L as K; MR 10 to 100 mg/L as K
    Documented instrumentsHI97750
    Reagent dose6 drops A and 1 packet B per determination in either range
    Sample and zero preparationLR: 10 mL sample; MR: dilute 20 mL to 100 mL with deionized water, then use 10 mL
    Timing on HI97750Shake gently 1 minute after B; wait 3 minutes; invert 5 times before final reading
    MatrixThe manual lists chloride above 12000 mg/L, calcium above 10000 mg/L as CaCO₃, magnesium above 8000 mg/L as CaCO₃, sodium above 8000 mg/L and ammonium above 10 mg/L as potential interferences. Check the sample matrix rather than assuming unrestricted brine or seawater suitability.
    Pack contentsA complete potassium set containing liquid reagent A and powder reagent B for 100 or 300 determinations. Both HI97750 ranges use 6 drops A and one B packet. HI93703-51 dispersing agent is not part of the documented HI97750 procedure; no additional dispersing step is required.
    Separate requirementsCompatible meter, cuvettes, stoppers, caps, method-required sample/dispensing accessories and CAL Check standards; volumetric flask and deionized water for MR dilution
    Shipping limitationDangerous Goods; ground shipment only

    Ordering Information

    Catalog numberComplete reagent pack
    HI93750-01100 tests — complete potassium reagent set
    HI93750-03300 tests — complete potassium reagent set

    Shipping, Return, and Warranty Policy

    At Blue Tiger Scientific, we are proud to distribute authentic Hanna Instruments products. The following policies apply to all purchases made through our store.

    Shipping Policy

    • All orders are shipped via FedEx from Hanna Instruments' fulfillment center in Smithfield, Rhode Island.
    • FedEx Ground is the standard method; expedited options (2-Day, Overnight) are available for an additional cost.
    • Hanna Instruments only ships to residential or business addresses within the United States (no PO Boxes).
    • Orders placed after business hours, weekends, or holidays will ship the next business day, pending product and carrier availability.
    • Shipping timelines may vary depending on product availability, order size, and carrier logistics.
    • Blue Tiger Scientific and Hanna Instruments are not responsible for carrier delays, weather-related disruptions, or other unforeseen delivery issues once the order has shipped.

    Return Policy

    • All return requests must be coordinated directly with Blue Tiger Scientific before a product is sent back.
    • Products must have a Return Merchandise Authorization (RMA) number issued by Blue Tiger Scientific before being returned.
    • Products must be returned in new or like-new condition, with the original packaging and proof of purchase (such as an invoice or packing slip).
      Customers are responsible for return shipping costs unless otherwise stated.
    • Unauthorized returns (without an RMA) will be refused and returned at the customer's expense.
      Returns are subject to inspection and approval. Products not in acceptable condition may be denied credit.
    • Some products, such as custom-built, special order, and hazardous materials, may not be eligible for return.

    Warranty Policy

    • Hanna Instruments' products are covered by their Limited Warranty against defects in materials and workmanship under normal use.
    • All warranty claims and services are handled directly by Hanna Instruments.
    • Warranty periods vary by product. If unspecified, the standard warranty term is 18 months from the Hanna Instruments factory ship date.
    • If you experience an issue believed to be covered under warranty, please contact Hanna Instruments for guidance on how to submit your warranty request to Hanna Instruments.
    • Customers must follow Hanna Instruments' warranty submission process, including providing a claim number, product model and serial number, and a detailed description of the issue.
    • Hanna Instruments will determine whether products are repaired, replaced, or otherwise serviced under warranty.
    • Note: Warranty coverage does not apply to damage caused by misuse, accidents, unauthorized repairs, or improper storage and handling.

    Important Notes

    • Some Hanna Instruments products are custom-made or built to order. Shipping will occur once the product is ready.
    • In the event of backorders, you may request partial shipments.
    • Delays beyond our control (e.g., material shortages, import delays, or carrier issues) may occasionally occur.

Questions about a product? Ask us here.

email us here: info@bluetigerscientific.com or live chat

Your Carts

Which would you like to open?

Shopping Cart

0 items

Review items for online purchase.

View Cart

Quote Cart

0 items

Review items and request a quote.

View Quote Cart

Your two carts stay separate. Nothing is moved or combined.