Hanna HI93720 Calcium Hardness LR Reagents, 100/300 Tests Photometer Reagents Hanna Instruments
Hanna HI93720 Calcium Hardness LR Reagents, 100/300 Tests Photometer Reagents Hanna Instruments

Hanna HI93720 Calcium Hardness LR Reagents, 100/300 Tests

$74.79

    Resolve low-level calcium hardness, not dissolved calcium. Keep HI97720 water-sample testing supplied with the complete matched calcium hardness lr reagent set.

    • HI93720-01: 100 tests; HI93720-03: 300 tests
    • HI97720: 0.00 to 2.70 mg/L as CaCO₃
    • Reagents only; meter, cuvettes and CAL Check standards separate

See All Specs

aeroplane.png__PID:119d4b1a-2623-400c-8fa4-aec5b49a3bf0

Fast Delivery

Most items ship same day

protection.png__PID:4b1a2623-e00c-4fa4-aec5-b49a3bf05f92

Manufacturer Warranty

All products are covered by manufacturer warranty

technical-support.png__PID:e00c8fa4-aec5-449a-bbf0-5f923431936b

After-Sale Support

24/7 support

HANNA PHOTOMETER REAGENTS · HI93720

Resolve low-level calcium hardness, not dissolved calcium.

Keep the HI97720 calcium-hardness method supplied with the matched three-liquid reagent set. Choose 100 or 300 determinations for low-hardness water monitoring without changing the chemistry or reporting basis.

Official Hanna HI93720 reagent set reference photo
Official Hanna calcium-hardness reagent-set reference photo. Printed lot and expiry markings are illustrative, not supplied-stock dates.

Match the method, not just the element name

HI97720: 0.00 to 2.70 mg/L as CaCO₃. Calmagite chemistry measures calcium hardness as calcium carbonate over 0.00 to 2.70 mg/L as CaCO₃ on the HI97720. This is not direct dissolved calcium as Ca²⁺, not HI937521 oxalate reagents, and not an undiluted swimming-pool hardness range.

What the selected pack supplies

A complete calcium-hardness A/B/C liquid reagent set for the selected 100 or 300 tests. This is reagent replenishment, not a photometer or CAL Check kit.

Where this method fits

The HI97720 is intended for low hardness levels such as those found in water-purification systems. For a documented 100-fold dilution, add 0.5 mL of sample to a clean, dry 50 mL beaker and bring to the 50 mL mark with hardness-free water. Account for dilution when interpreting the original sample; do not treat the native range as a pool-water range.

The HI97720 workflow at a glance

Rinse a graduated beaker with unreacted sample and fill it to 50 mL. Add 0.5 mL of reagent A, swirl, then add 0.5 mL of reagent B and swirl. Fill two cuvettes with 10 mL of this treated sample. Add one drop of reagent C to the first cuvette only, stopper, cap and invert to mix: this is the blank. Align its cap notch with the holder groove and zero the HI97720. Replace it with the second cuvette, aligned in the same way, and press Read for calcium hardness as CaCO₃.

Check the sample matrix

Excessive heavy metals can interfere. Calcium contamination in the beaker, syringes or cuvettes affects this low-level test; repeat cleanliness checks until results are consistent. Dilutions require hardness-free water, not ordinary tap water.

Choose a pack around your workload

HI93720-01 supplies 100 tests; HI93720-03 supplies 300 tests. The larger pack supplies three times the determinations, not a different measuring range. One order quantity means one complete selected pack. Choose by testing frequency and the lot/expiry markings on supplied stock.

What to have beside the reagent set

Supply the compatible photometer, cuvettes with stoppers and caps, and the measuring/dispensing accessories required by the method separately. Use hardness-free water for calcium-hardness dilution. CAL Check standards are separate instrument-verification materials, not replacements for sample reagents.

Handling before you begin

Review all three US component SDS. Reagent B may corrode metals and causes severe skin burns and eye damage; use the specified protection and handling controls. Hanna flags both complete pack sizes as Dangerous Goods for ground shipment only.

Use the linked HI97720 instrument manual as the method reference for the complete procedure, precautions and instrument operation. This is not a separate reagent instruction manual.

Frequently Asked Questions

Should I choose 100 or 300 tests?

HI93720-01 supplies 100 tests; HI93720-03 supplies 300 tests. Both are complete reagent sets for the same method. Choose by expected workload and marked expiry dates, not by a different analytical range.

Is this calcium hardness or dissolved calcium?

It is calcium hardness, reported as mg/L CaCO₃. HI97720 uses calmagite chemistry over 0.00 to 2.70 mg/L as CaCO₃. HI937521 measures calcium directly with different chemistry and is not a substitute.

Why does the method need two cuvettes?

Both receive 10 mL of the sample treated with A and B. Add one drop of C to only the first cuvette and use that as the blank. Zero with the blank, then read the second cuvette. Do not add C to both.

Can I measure pool water without dilution?

Do not assume that. The native HI97720 range is low-level hardness. The manual documents a 100-fold dilution using 0.5 mL sample brought to 50 mL with hardness-free water; follow it and account for dilution when interpreting results.

Why is clean glassware especially important?

This low-range method detects calcium contamination from the beaker, syringes and sample cells. Repeat the cleanliness test until results are consistent and use hardness-free water for dilution.

Which photometers are documented for HI93720?

Both packs are documented for HI97720. The procedure and range on this page use HI97720 as the primary reference; follow the exact instrument method when using another model.

Which sample conditions should I check first?

Excessive heavy metals can interfere. Calcium contamination in the beaker, syringes or cuvettes affects this low-level test; repeat cleanliness checks until results are consistent. Dilutions require hardness-free water, not ordinary tap water.

Are the instrument, cuvettes and CAL Check standards included?

No. These are reagent-only packs. Provide the photometer, cuvettes with stoppers and caps, and the method-required sampling and dispensing accessories separately. CAL Check standards are separate instrument-validation materials, not test reagents.

    Product Specifications

    SpecificationDetail
    ParameterCalcium Hardness LR
    Reagent formatLiquid calcium-hardness reagents A (HI93720A-0), B (HI93720B-0) and C (HI93720C-0)
    MethodAdaptation of Standard Methods for the Examination of Water and Wastewater, 18th Edition, Calmagite Method
    Primary instrumentHI97720
    Range and reporting basis0.00 to 2.70 mg/L as CaCO₃
    Documented instrumentsHI97720
    Reagent dose0.5 mL A and 0.5 mL B into 50 mL sample; 1 drop C into the 10 mL blank only
    Sample and zero preparation50 mL treated sample split into two 10 mL cuvettes; C-treated cuvette is the blank
    Timing on HI97720Zero the C-treated blank; read the second cuvette
    MatrixExcessive heavy metals can interfere. Calcium contamination in the beaker, syringes or cuvettes affects this low-level test; repeat cleanliness checks until results are consistent. Dilutions require hardness-free water, not ordinary tap water.
    Pack contentsA complete calcium-hardness A/B/C liquid reagent set for the selected 100 or 300 tests. This is reagent replenishment, not a photometer or CAL Check kit.
    Separate requirementsCompatible meter, cuvettes, stoppers, caps, method-required sample/dispensing accessories and CAL Check standards; hardness-free water for dilution
    Shipping limitationDangerous Goods; ground shipment only

    Ordering Information

    Catalog numberComplete reagent pack
    HI93720-01100 tests — complete calcium hardness lr reagent set
    HI93720-03300 tests — complete calcium hardness lr reagent set

    Shipping, Return, and Warranty Policy

    At Blue Tiger Scientific, we are proud to distribute authentic Hanna Instruments products. The following policies apply to all purchases made through our store.

    Shipping Policy

    • All orders are shipped via FedEx from Hanna Instruments' fulfillment center in Smithfield, Rhode Island.
    • FedEx Ground is the standard method; expedited options (2-Day, Overnight) are available for an additional cost.
    • Hanna Instruments only ships to residential or business addresses within the United States (no PO Boxes).
    • Orders placed after business hours, weekends, or holidays will ship the next business day, pending product and carrier availability.
    • Shipping timelines may vary depending on product availability, order size, and carrier logistics.
    • Blue Tiger Scientific and Hanna Instruments are not responsible for carrier delays, weather-related disruptions, or other unforeseen delivery issues once the order has shipped.

    Return Policy

    • All return requests must be coordinated directly with Blue Tiger Scientific before a product is sent back.
    • Products must have a Return Merchandise Authorization (RMA) number issued by Blue Tiger Scientific before being returned.
    • Products must be returned in new or like-new condition, with the original packaging and proof of purchase (such as an invoice or packing slip).
      Customers are responsible for return shipping costs unless otherwise stated.
    • Unauthorized returns (without an RMA) will be refused and returned at the customer's expense.
      Returns are subject to inspection and approval. Products not in acceptable condition may be denied credit.
    • Some products, such as custom-built, special order, and hazardous materials, may not be eligible for return.

    Warranty Policy

    • Hanna Instruments' products are covered by their Limited Warranty against defects in materials and workmanship under normal use.
    • All warranty claims and services are handled directly by Hanna Instruments.
    • Warranty periods vary by product. If unspecified, the standard warranty term is 18 months from the Hanna Instruments factory ship date.
    • If you experience an issue believed to be covered under warranty, please contact Hanna Instruments for guidance on how to submit your warranty request to Hanna Instruments.
    • Customers must follow Hanna Instruments' warranty submission process, including providing a claim number, product model and serial number, and a detailed description of the issue.
    • Hanna Instruments will determine whether products are repaired, replaced, or otherwise serviced under warranty.
    • Note: Warranty coverage does not apply to damage caused by misuse, accidents, unauthorized repairs, or improper storage and handling.

    Important Notes

    • Some Hanna Instruments products are custom-made or built to order. Shipping will occur once the product is ready.
    • In the event of backorders, you may request partial shipments.
    • Delays beyond our control (e.g., material shortages, import delays, or carrier issues) may occasionally occur.

Questions about a product? Ask us here.

email us here: info@bluetigerscientific.com or live chat

Your Carts

Which would you like to open?

Shopping Cart

0 items

Review items for online purchase.

View Cart

Quote Cart

0 items

Review items and request a quote.

View Quote Cart

Your two carts stay separate. Nothing is moved or combined.