Match the water activity (aw) specified in your quality-control procedure to the catalog number below. Each row is a separately orderable saturated salt standard for the VEE GEE 47002 Water Activity Meter. The meter is not included.

47011 is magnesium chloride at 0.328 aw; 47014 is potassium chloride at 0.843 aw. Neither replaces the 0.753 aw sodium chloride standard (47013). Record both the salt and the catalog number when reordering.
Compare the standard under the same conditions as the sample, including temperature and meter settings. Water activity is temperature dependent. Heathrow recommends considering calibration when the difference from the reference exceeds 0.02 aw; this is a check on the meter, not an accuracy tolerance for the salt.
No. Select a catalog number to order that salt and reference point. The seven options are not a calibration set. Heathrow does not specify an inner-container count in its ordering table.
VEE GEE lists all seven as optional calibration salts for its 47002 Water Activity Meter. Neither the instrument nor a sensor is included with a salt order.
47012, magnesium nitrate, is listed at 0.529 aw. For comparison, the next lower point is 47011 at 0.328 aw; the next higher point is 47013 at 0.753 aw.
47015 is ammonium dihydrogen phosphate at 0.930 aw. Choose 47016, potassium sulfate, if your procedure calls for 0.973 aw instead.
No. Water activity is a ratio, not a percentage. At equilibrium and at the same temperature, 0.753 aw corresponds to 75.3% equilibrium relative humidity. It is not the room humidity you need to set.
Use the reference applicable to your method and keep the standard and sample under matching conditions. A published saturated-salt reference table for these seven chemistries gives the listed values at 25°C. Heathrow does not provide a temperature-correction table or a lot-specific certificate in its ordering chart. Do not assume the same values at other temperatures.
Heathrow describes the standards as reusable and recommends replacement after one year. Do not puncture the waterproof, breathable film. Inspect the container and follow the handling instructions supplied with the standard.