HANNA PHOTOMETER REAGENTS · HI93706
Keep phosphorus results on the P basis.
Replenish the paired liquid-and-powder chemistry for HI97706 phosphorus measurements. Select 100 or 300 tests for your routine without confusing phosphorus as P with a different phosphate reagent kit or reporting scale.
Official Hanna phosphorus-page generic packet reference image, not a photograph of the complete liquid A/powder B set. Printed expiry is illustrative, not the expiry of supplied stock.
Match the method, not just the element name
HI97706: 0.0 to 15.0 mg/L as P. The amino acid method produces a blue tint from orthophosphate and reports the phosphorus portion. The HI97706 range is 0.0 to 15.0 mg/L as P. A result expressed as phosphate (PO₄) is not numerically equivalent; changing the displayed chemical form does not make this a different reagent chemistry.
What the selected pack supplies
A complete matched phosphorus set with liquid reagent A (HI93706A-0) and powder reagent B (HI93706B-0) for the selected 100 or 300 tests. Per-test doses describe the procedure, not independently verified retail bottle counts, bottle fills or packet inventory.
Where this method fits
Use where the testing program calls for this orthophosphate-based phosphorus method. This refill is not a digestion kit for total or acid-hydrolyzable phosphorus. Keep the sample treatment and reporting basis specified by your procedure; do not exchange it with a phosphate kit solely because the element name looks similar.
The HI97706 workflow at a glance
Fill a clean cuvette with 10 mL of unreacted sample. Replace the stopper and cap, align the cap notch with the holder groove and zero the HI97706. Remove the cuvette, add 10 drops of HI93706A-0 and one packet of HI93706B-0, then stopper, cap and shake gently until completely dissolved. Return it to the holder with the same alignment and use the 5-minute Read countdown. Record the result in mg/L as P unless you deliberately select and label another chemical form.
Check the sample matrix
Sulfide may interfere. Listed thresholds are chloride above 150000 mg/L, magnesium above 40000 mg/L as CaCO₃, calcium above 10000 mg/L as CaCO₃ and ferrous iron above 100 mg/L. Check these matrix conditions rather than assuming universal water-sample suitability.
Choose a pack around your workload
HI93706-01 supplies 100 tests; HI93706-03 supplies 300 tests. The larger pack supplies three times the determinations, not a different measuring range. One order quantity means one complete selected pack. Choose by testing frequency and the lot/expiry markings on supplied stock.
What to have beside the reagent set
Supply the compatible photometer, cuvettes with stoppers and caps, and the measuring/dispensing accessories required by the method separately. Provide scissors to open the powder packet. CAL Check standards are separate instrument-verification materials, not replacements for sample reagents.
Handling before you begin
Review both US component SDS. Liquid reagent A may corrode metals and causes severe skin burns and eye damage; powder reagent B causes serious eye damage. Use the specified protection, handling and disposal controls. Hanna identifies both complete pack sizes as Dangerous Goods for ground shipment only.
Use the linked HI97706 instrument manual as the method reference for the complete procedure, precautions and instrument operation. This is not a separate reagent instruction manual.
Frequently Asked Questions
Should I choose 100 or 300 tests?
HI93706-01 supplies 100 tests; HI93706-03 supplies 300 tests. Both are complete reagent sets for the same method. Choose by expected workload and marked expiry dates, not by a different analytical range.
Does this report phosphorus or phosphate?
The quoted HI97706 range is 0.0 to 15.0 mg/L as phosphorus, P, from orthophosphate. The meter can display converted PO₄ or P₂O₅ chemical forms, but those values are not numerically the same. Label the basis with the result.
Can I use this to determine total phosphorus?
Not as a direct total-phosphorus or digestion method. The documented amino acid reaction measures the phosphorus portion of orthophosphate. This set does not establish a digestion capability for total or acid-hydrolyzable phosphorus.
Which reagent components are needed?
The HI97706 method uses liquid HI93706A-0 reagent A and powder HI93706B-0 reagent B: 10 drops A plus one B packet for 10 mL sample. Both are part of the matched chemistry; do not substitute a different phosphate kit.
When do I zero and read the HI97706?
Zero with 10 mL unreacted sample before adding A or B. Add 10 drops A and one packet B, shake gently until dissolved, and use the 5-minute Read countdown. The linked instrument manual gives the complete procedure.
Which photometers are documented for HI93706?
Both packs are documented for HI97706. The procedure and range on this page use HI97706 as the primary reference; follow the exact instrument method when using another model.
Which sample conditions should I check first?
Sulfide may interfere. Listed thresholds are chloride above 150000 mg/L, magnesium above 40000 mg/L as CaCO₃, calcium above 10000 mg/L as CaCO₃ and ferrous iron above 100 mg/L. Check these matrix conditions rather than assuming universal water-sample suitability.
Are the instrument, cuvettes and CAL Check standards included?
No. These are reagent-only packs. Provide the photometer, cuvettes with stoppers and caps, and the method-required sampling and dispensing accessories separately. CAL Check standards are separate instrument-validation materials, not test reagents.