APERA INSTRUMENTS / PION400-CL / AI4388
The Apera PION400-CL is a portable pH and chloride ion meter kit for chloride-ion analysis in prepared aqueous samples and process-water investigations. Its dedicated Cl⁻ electrode supports ion-selective measurement while the IP57 meter brings a backlit display and guided calibration to bench or field work.
The chloride electrode is specified for 1.775–3550 mg/L, sample pH 2–11 and 5–60°C. Choose PION400 when the same workflow also needs pH; choose ION400 when ion concentration, mV and temperature cover the task.
Choose the ion you report, then decide whether you also need pH. Each ion requires its own sensor, calibration standards and sample-conditioning method.
Linked cards open current store listings. Other cards compare configurations; they do not imply current storefront availability.
The platform supports ion-selective testing with the corresponding electrode and calibration. Changing an ion setting does not turn one selective membrane into a sensor for every ion.
PION400 portable meter and the chloride ion-selective electrode configuration identified by manufacturer part number AI4388. This PION400 configuration adds pH measurement capability. Calibration solution concentrations, bottle sizes and the complete accessory pack should be confirmed for the ordered configuration; the family photo does not show an itemized kit.
The dedicated solid-state membrane responds selectively to chloride in a prepared aqueous sample. The combination design brings measuring and reference functions together; temperature support must match the meter.
Condition and calibrate the chloride electrode with suitable standards before measurement. Keep samples within 2–11 pH and 5–60°C; apply the method-specific ionic-strength adjustment and interference controls. Measures chloride ions, not free chlorine, total chlorine or chlorine disinfectant residual.
This chloride version uses a solid-state membrane. The manufacturer classification separates PVC, solid-state and crystalline sensors because different ions call for different sensing materials. PVC durability and storage claims should not be transferred to every membrane type.
The handbook diagram distinguishes IndSen 2071M analog combination electrodes from 2071S digital models and 1121M measuring electrodes. The analog 2071M branch is the relevant family here; the other architectures are shown for comparison.
Match the exact electrode connection and temperature support to the meter. A shared body shape or plug appearance is not sufficient to establish interchangeability.
Calibrate with suitable ion standards, then prepare standards and samples using the same method. The 400-series ion mode supports one- to three-point manual calibration; select appropriate points for the samples you expect.
Ionic strength, interfering ions, temperature, sample pH and deposits can affect the reading. Apera’s handbook checklist shows why electrode selection and sample preparation belong in the same workflow.
Follow the instructions for this ion-selective electrode when conditioning, cleaning and storing it. Use appropriate standards to check response after maintenance. The measuring membrane and reference junction need contact with the prepared sample during measurement.
The PVC family uses solid-electrolyte technology; this does not remove the need for conditioning and calibration. Use model-specific care instructions rather than applying a general pH-probe soaking routine to every ISE.
PION400 adds a pH channel to ion, mV and temperature measurement. With a suitable pH/temperature electrode, the meter supports 0.00–14.00 pH at 0.01 pH resolution and ±0.01 pH ±1 digit, with one- to three-point automatic pH calibration.
Keep the pH electrode’s calibration and care routine separate from the ion electrode’s method. Review the exact electrode and buffer pack for the ordered configuration.
ION400-K potassium kit — compare the ion-only meter format PION400-K potassium/pH kit — compare the pH-capable format 601 stirrer with electrode holder — support a consistent sample routine
It measures chloride ions (Cl⁻) with the corresponding electrode, plus mV and temperature. PION400 also provides a pH mode with a suitable pH electrode.
Apera’s electrode-family catalog specifies 1.775–3550 mg/L. The usable sample range also depends on calibration, interferences and sample preparation.
The electrode specification calls for 5–60°C and a sample pH of 2–11. These limits are narrower than the meter’s general temperature display range.
Measures chloride ions, not free chlorine, total chlorine or chlorine disinfectant residual.
Use suitable chloride standards and the meter’s 1–3 point manual ion calibration. Match the sample-conditioning procedure between standards and samples.
No. The chloride electrode is selective for Cl⁻. Measuring another supported ion requires the corresponding electrode, standards and calibration.
ION400 focuses on ion concentration, mV and temperature. PION400 adds pH measurement capability; choose it when your workflow also calls for pH testing.
The meter has an IP57 waterproof and dustproof rating and uses four AA batteries. Protect connections and follow the operating instructions; the rating does not remove sample and electrode operating limits.
Review PION400-CL AI4388 against your measurement method, sample limits and host connection. Return to the price and ordering controls when the configuration fits your work.
Manufacturer warranty coverage is subject to Apera’s terms for this exact product. Contact Blue Tiger Scientific for help confirming coverage or submitting a manufacturer warranty inquiry.
For technical support, provide the manufacturer part number, meter model and a description of the issue. Blue Tiger Scientific can help coordinate product selection and manufacturer support.
For a return request, contact Blue Tiger Scientific before sending the product back. Eligibility and any authorization requirements follow the applicable store return policy; manufacturer warranty service and store returns are separate processes.
email us here: info@bluetigerscientific.com or live chat
Shipping and taxes calculated at checkout. Your quote list is separate.
Review your items, then tell Blue Tiger Scientific where your order is going. Your shopping cart stays unchanged.
International enquiries are welcome. Pricing, shipping availability, freight and any import requirements are confirmed manually.
Prices shown are for reference. We’ll confirm pricing, availability and shipping in your quote.
For eligible Accucold / Summit equipment. White glove and inside delivery are not included and are quoted separately. Shipping availability and final freight are confirmed with your quote, including Alaska, Hawaii and mixed-product requests.
Optional preferences only. Terms must be confirmed by Blue Tiger Scientific; no payment is collected and no terms are approved by this request.
Loading the security check…
* Required. Contact details and eligible quote-cart items will be submitted to Blue Tiger Scientific for this enquiry. No payment is collected.
Which would you like to open?
Review items for online purchase.
Review items and request a quote.
Your two carts stay separate. Nothing is moved or combined.